Spacelabs Healthcare Xprezzon Beside MONITOR(9913930)
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K112962, from Spacelabs Medical, Inc. (Issaquah, WA). Product code MHX, class 2.
As of Nov 17, 2026, K112962 is an FDA 510(k) held by Spacelabs Medical, Inc., substantially equivalent on Nov 2, 2011 after 28 days, with 7 recalls tied to it in FDA's recall database.
- Substantially equivalent: Nov 2, 2011
- Review time: 28 days
- Recalls: 7
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Spacelabs Healthcare Xprezzon Bedside Monitor passively displays data generated by Spacelabs parameter modules, Flexport interfaces, and other Spacelabs SDLC based products as waveform and numeric displays, trends and alarms.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Spacelabs Medical Patient MonitorsK102422 · Spacelabs Medical, Inc. · Dec 7, 2010
Recalls
Recalls FDA ties to K112962, newest first; each opens FDA's recall record| Sep 5, 2025 | Z-0167-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| Sep 5, 2025 | Z-0169-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| Sep 5, 2025 | Z-0168-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of… | Open, Classified |
| Jan 28, 2015 | Z-1144-2015Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings… | Terminated |
| Oct 5, 2012 | Z-0196-2013Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation. | Terminated |
| Jun 12, 2012 | Z-2019-2012There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters… | Terminated |
| May 22, 2012 | Z-1705-2012The screw used to connect the external power supply to the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, may damage an internal capacitor. The damage may result in an inability of the monitor to power on,… | Terminated |
Other MHX Clearances
The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- M12 Telemetry SystemK252840 · Global Instrumentation, LLC · May 29, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 5, 2011
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files