Arca
Skip to content

Spacelabs Healthcare Xprezzon Beside MONITOR(9913930)

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K112962, from Spacelabs Medical, Inc. (Issaquah, WA). Product code MHX, class 2.

As of Nov 17, 2026, K112962 is an FDA 510(k) held by Spacelabs Medical, Inc., substantially equivalent on Nov 2, 2011 after 28 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 2, 2011
  • Review time: 28 days
  • Recalls: 7

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Spacelabs Healthcare Xprezzon Bedside Monitor passively displays data generated by Spacelabs parameter modules, Flexport interfaces, and other Spacelabs SDLC based products as waveform and numeric displays, trends and alarms.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K112962, newest first; each opens FDA's recall record
Sep 5, 2025Z-0167-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…Open, Classified
Sep 5, 2025Z-0169-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…Open, Classified
Sep 5, 2025Z-0168-2026Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of…Open, Classified
Jan 28, 2015Z-1144-2015Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings…Terminated
Oct 5, 2012Z-0196-2013Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.Terminated
Jun 12, 2012Z-2019-2012There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters…Terminated
May 22, 2012Z-1705-2012The screw used to connect the external power supply to the Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, may damage an internal capacitor. The damage may result in an inability of the monitor to power on,…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Special
Received
Oct 5, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.