Carescape Monitor B650

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K102239, from GE Healthcare Finland Oy (Helsinki, FI). Product code MHX, class 2.

As of Nov 17, 2026, K102239 is an FDA 510(k) held by GE Healthcare Finland Oy, substantially equivalent on Oct 18, 2010 after 70 days, with 8 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 18, 2010
  • Review time: 70 days
  • Recalls: 8
  • Cited as a predicate by: 6 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

6 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102239, newest first; each opens FDA's recall record
May 25, 2018Z-2340-2018When multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be…Open, Classified
May 4, 2015Z-1606-2015The heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message…Terminated
Jul 7, 2014Z-2114-2014Possible sound loss associated with the CARESCAPE Monitor B650. When an external display is connected to the CARESCAPE Monitor B650 using a USB extender cable with an active USB hub, the system may experience a sound…Terminated
Feb 28, 2014Z-1175-2014GE Healthcare has recently become aware of potential safety issues associated with the CARESCAPE Monitor B850 and CARESCAPE Monitor B650. These issues include both NIBP (Non-Invasive Blood Pressure) and Bed to Bed…Terminated
Dec 27, 2013Z-0854-2014GE Healthcare has recently become aware of potential safety issues associated with the CARESCAPE Monitor B850 and CARESCAPE Monitor B650. These included 25 issues related to NIBP (Non-InvasiVentilation Alarms:ve Blood…Terminated
Jul 12, 2012Z-2256-2013GE Healthcare has recently become aware of a number of potential issues associated with the CARESCAPE Monitor B650. No injuries or illnesses have been reported.Terminated
Sep 30, 2011Z-0823-2013GE Healthcare has become aware of potential issues associated with the CARESCAPE Monitor B650. There have been no reported illnesses or injuries that have occurred with the use of the device.Terminated
Mar 23, 2011Z-1930-2011GE Healthcare has become aware of a potential safety issue associated with the CARESCAPE Monitor B650 & B850 Cardiac Output measurement using Patient Data Module (PDM), and activation of ECG Leads Off alarm in specific…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Traditional
Received
Aug 9, 2010
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files