Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K131223, from GE Healthcare Finland Oy (Helsinki, FI). Product code MHX, class 2.

As of Nov 17, 2026, K131223 is an FDA 510(k) held by GE Healthcare Finland Oy, substantially equivalent on Aug 28, 2013 after 120 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 28, 2013
  • Review time: 120 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The CARESCAPE Monitor B650 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K131223, newest first; each opens FDA's recall record
May 25, 2018Z-2340-2018When multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be…Open, Classified
May 4, 2015Z-1606-2015The heart rate could be calculated from pacer pulses without indication that pacemaker detection is OFF in some combinations of Bx50 monitors, a PDM, and CIC/CSCS. Then the monitor does not show the Pacer Off message…Terminated
Jun 13, 2014Z-2085-2014Partial loss of monitoring with the CARESCAPE Monitor B850, B650 or B450. When connecting the PDM (Patient Data Module) to the CARESCAPE B850, B650 or B450 monitor with software version 2 the PDM may not start…Terminated
Feb 28, 2014Z-1175-2014GE Healthcare has recently become aware of potential safety issues associated with the CARESCAPE Monitor B850 and CARESCAPE Monitor B650. These issues include both NIBP (Non-Invasive Blood Pressure) and Bed to Bed…Terminated

Other MHX Clearances

The newest 510(k)s for the same product code; their median review was 226 days; see every MHX device

Record

Type
Traditional
Received
Apr 30, 2013
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files