Rosa Surgical Device
Neurological Stereotaxic Instrument. 510(k) K101791, from Medtech S.A.S (Montpellier, FR). Product code HAW, class 2.
As of Nov 17, 2026, K101791 is an FDA 510(k) held by Medtech S.A.S, substantially equivalent on Sep 23, 2010 after 87 days, with 12 recalls tied to it in FDA's recall database.
- Substantially equivalent: Sep 23, 2010
- Review time: 87 days
- Recalls: 12
- Cited as a predicate by: 4 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Rosa Surgical Device, Model Rosa 1.1K092239 · Medtech S.A · Nov 17, 2009
- Steathstation Treatment Guidance PlatformK001801 · Medtronic Surgical Navigation · Jun 30, 2000
- Vectorvision Cranial/EntK023651 · Brainlab AG · Feb 17, 2004
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Mazor XK172522 · Mazor Robotics , Ltd. · Sep 15, 2017
- ROSA SpineK151511 · Medtech S.A · Jan 4, 2016
- ROSA BrainK151359 · Medtech S.A · Dec 18, 2015
- SYMBIS Surgical SystemK143420 · Imris, Inc. · Oct 30, 2015
Recalls
Recalls FDA ties to K101791, newest first; each opens FDA's recall record| May 8, 2018 | Z-2108-2018Replacement of units lacking an updated device approval. | Terminated |
| May 27, 2016 | Z-0105-2018Insufficient amount of washers in the Telescopic Arm, allowing some movement of the device. | Terminated |
| Apr 8, 2016 | Z-0115-2018Potential for software to change the final tool orientation for the command position without command. | Terminated |
| Jan 20, 2016 | Z-3117-2017Complaints were received reporting the system would freeze/shut down while in Fulgurate mode. | Terminated |
| Nov 6, 2014 | Z-0160-2018Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended). | Terminated |
| Sep 5, 2014 | Z-0184-2018The software issue described was corrected in the modification to the MXTTOUT controller parameter settings. | Terminated |
| Jun 4, 2014 | Z-0161-2018Software corrections reactivating the cooperative endoscopy mode. | Terminated |
| Feb 14, 2014 | Z-0102-2018An undetected shift between the information displayed in the navigation software and the actual patient anatomy | Terminated |
| Dec 18, 2013 | Z-0096-2018Possible break in the connector of the Force Sensor. | Terminated |
| Feb 10, 2013 | Z-0178-2018Issue with optional neuro-endoscopy module detected under specific conditions during internal testing. | Terminated |
| Feb 5, 2013 | Z-0099-2018Observed instability of the device. | Terminated |
| Apr 19, 2012 | Z-0100-2018Potential failure of the optical distance sensor due to cable disconnection. | Terminated |
Other HAW Clearances
The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device- NeuroAlignK261997 · Medivis, Inc. · Oct 26, 2026
- Disposable Reflective Marker Spheres (AM-BK115S, AM-BL130S)K262214 · Guangzhou Aimooe Technology Co., Ltd. · Oct 15, 2026
- Geniant Cranial (Navigated Neurosurgical Positioning Robot)K253856 · Koh Young Technology, Inc. · Sep 10, 2026
- Leksell® Vantage Stereotactic SystemK261932 · Elekta Solutions AB · Aug 5, 2026
- Zeta Navigation SystemK253663 · Zeta Surgical, Inc. · May 1, 2026
- Stealth AXiS Cranial clinical applicationK253379 · Medtronic Navigation, Inc. · Mar 26, 2026
- NeuroAlign softwareK231897 · Medivis, Inc. · Oct 21, 2025
- SOLOPASS 2.0 SystemK251317 · Intravent Medical Partners · Sep 5, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 28, 2010
- Review panel
- Neurology
- Regulation
- 21 CFR 882.4560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files