Neurological Stereotaxic Instrument. 510(k) K151359, from Medtech S.A (Montpellier, FR). Product code HAW, class 2.

As of Nov 17, 2026, K151359 is an FDA 510(k) held by Medtech S.A, substantially equivalent on Dec 18, 2015 after 212 days, with 7 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Dec 18, 2015
  • Review time: 212 days
  • Recalls: 7
  • Cited as a predicate by: 2 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K151359, newest first; each opens FDA's recall record
Jan 6, 2020Z-1151-2020Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the…Terminated
Jan 6, 2020Z-1150-2020Some cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the…Terminated
Sep 10, 2019Z-0172-2020The instrument holder may be sent on a trajectory that is not within the intended target. If it is not corrected, the associated device may be placed incorrectly.Terminated
Feb 27, 2017Z-1754-2017Unapproved change made by the supplier.Terminated
Oct 11, 2016Z-2868-2017Complaint of the head holder connector locking up mechanically when tightened.Terminated
Jul 29, 2016Z-0116-2018Non-conforming product label.Terminated
Jul 20, 2016Z-0167-2018Communication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.Terminated

Other HAW Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Traditional
Received
May 20, 2015
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.