Arca
Skip to content

Rosa Surgical Device, Model Rosa 1.1

Neurological Stereotaxic Instrument. 510(k) K092239, from Medtech S.A (Montpellier, FR). Product code HAW, class 2.

As of Nov 17, 2026, K092239 is an FDA 510(k) held by Medtech S.A, substantially equivalent on Nov 17, 2009 after 117 days, with 9 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 17, 2009
  • Review time: 117 days
  • Recalls: 9
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers registration.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K092239, newest first; each opens FDA's recall record
Jan 20, 2016Z-3117-2017Complaints were received reporting the system would freeze/shut down while in Fulgurate mode.Terminated
Nov 6, 2014Z-0160-2018Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).Terminated
Sep 5, 2014Z-0184-2018The software issue described was corrected in the modification to the MXTTOUT controller parameter settings.Terminated
Jun 4, 2014Z-0161-2018Software corrections reactivating the cooperative endoscopy mode.Terminated
Feb 14, 2014Z-0102-2018An undetected shift between the information displayed in the navigation software and the actual patient anatomyTerminated
Dec 18, 2013Z-0096-2018Possible break in the connector of the Force Sensor.Terminated
Feb 10, 2013Z-0178-2018Issue with optional neuro-endoscopy module detected under specific conditions during internal testing.Terminated
Feb 5, 2013Z-0099-2018Observed instability of the device.Terminated
Apr 19, 2012Z-0100-2018Potential failure of the optical distance sensor due to cable disconnection.Terminated

Other HAW Clearances

The newest 510(k)s for the same product code; their median review was 162 days; see every HAW device

Record

Type
Traditional
Received
Jul 23, 2009
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.