Accelerator Aps
Electrode, Ion Specific, Potassium. 510(k) K093318, from Abbott Laboratories (Irving, TX). Product code CEM, class 2.
As of Nov 18, 2026, K093318 is an FDA 510(k) held by Abbott Laboratories, substantially equivalent on Feb 19, 2010 after 119 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 119 days
- Recalls: 2
- Cited as a predicate by: 3 devices
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Aeroset SystemK980367 · Abbott Laboratories · Apr 1, 1998
- Advia Centaur System With Streamlab Analytical WorkcellK082638 · Siemens Healthcare Diagnostics, Inc. · Feb 26, 2009
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- GLP systems TrackK213486 · Abbott Laboratories · Mar 10, 2022
- Flexlab 3.6, Accelerator A3600K121012 · Inpeco S.P.A. · Aug 31, 2012
- Fastep Hcg Pregancy Serum/Urine Cassette and Dipstick TestsK112101 · Polymed Therapeutics, Inc. · Jul 17, 2012
Recalls
Recalls FDA ties to K093318, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 2, 2020 | Z-2251-2020In the case of an AQM clot detection error a portion of the distilled water from the Aliquoter Module hydraulic circuit may be dispensed into the Primary Tube leading to sample dilution, which could lead to incorrect… | Terminated |
| Sep 7, 2018 | Z-2755-2019The firm identified customer sites where modules which the safety pacemaker label is missing. Information regarding the label included in the Operations Manual has been updated to specify: 1) the risk of malfunction… | Terminated |
Other CEM clearances
The newest 510(k)s for the same product code; their median review was 273 days; see every CEM device- ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutionsK253491 · Roche Diagnostics · Feb 12, 2026
- ABL90 FLEX PLUS SystemK241037 · Radiometer Medicals Aps · Jan 14, 2025
- EasyStat 300K220396 · Medica Corporation · Sep 26, 2023
- ISE ElectrodesK230890 · Randox Laboratories, Ltd. · Sep 8, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 23, 2009
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files