Flexlab 3.6, Accelerator A3600
Electrode, Ion Specific, Potassium. 510(k) K121012, from Inpeco S.P.A. (Segrate, Milan, IT). Product code CEM, class 2.
As of Nov 18, 2026, K121012 is an FDA 510(k) held by Inpeco S.P.A., substantially equivalent on Aug 31, 2012 after 150 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 150 days
- Recalls: 25+
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that performs a variety of instrument specific assays such as Sodium, Potassium and Chloride.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Accelerator ApsK093318 · Abbott Laboratories · Feb 19, 2010
Recalls
Recalls FDA ties to K121012, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Oct 30, 2025 | Z-0595-2026The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head… | Open |
| Oct 30, 2025 | Z-0596-2026The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head… | Open |
| Sep 3, 2025 | Z-0107-2026The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the… | Open |
| Sep 3, 2025 | Z-0106-2026The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the… | Open |
| Nov 3, 2023 | Z-0560-2024Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing. | Open |
| Mar 10, 2023 | Z-1560-2023, Z-1561-2023, Z-1559-2023, Z-1563-2023, Z-1562-2023Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either… | Open |
| Sep 10, 2021 | Z-0087-2022The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient… | Open |
| Aug 7, 2020 | Z-2975-2020, Z-2973-2020, Z-2974-2020Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results | Terminated |
| Jul 14, 2020 | Z-0433-2021, Z-0434-2021, Z-0432-2021When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with… | Terminated |
| Apr 7, 2020 | Z-1888-2020, Z-1887-2020, Z-1889-2020The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge. | Open |
| Apr 7, 2020 | Z-2198-2020Potential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the… | Open |
| Apr 7, 2020 | Z-2200-2020, Z-2199-2020In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary… | Open |
| Apr 7, 2020 | Z-2140-2020The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can… | Terminated |
| Apr 7, 2020 | Z-2114-2020The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design. | Terminated |
Other CEM clearances
The newest 510(k)s for the same product code; their median review was 273 days; see every CEM device- ISE indirect K for Gen.2; ISE indirect Na for Gen.2; ISE indirect Cl for Gen.2; cobas pro integrated solutionsK253491 · Roche Diagnostics · Feb 12, 2026
- ABL90 FLEX PLUS SystemK241037 · Radiometer Medicals Aps · Jan 14, 2025
- EasyStat 300K220396 · Medica Corporation · Sep 26, 2023
- ISE ElectrodesK230890 · Randox Laboratories, Ltd. · Sep 8, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Apr 3, 2012
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1600
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files