Aeroset System
Electrode, Ion-Specific, Chloride. 510(k) K980367, from Abbott Laboratories (Irving, TX). Product code CGZ, class 2.
As of Nov 18, 2026, K980367 is an FDA 510(k) held by Abbott Laboratories, substantially equivalent on Apr 1, 1998 after 62 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 62 days
- Recalls: 25+
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFPer 21 CFR-682.2160, the AEROSET System is a discrete photometric chemistry Analyzer for clinical use. The device is intended to duplicate manual analytical procedures by performing various steps such as pipetting, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Hitachi 717 Chemistry AnalyzerK872494 · Boehringer Mannheim Corp. · Jul 9, 1987
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- Alinity c ICT Sample DiluentK170320 · Abbott Laboratories · Oct 24, 2017
- Alinity C Glucose Reagent Kit, Alinity C SystemK170316 · Abbott Laboratories · Oct 19, 2017
- Accelerator ApsK093318 · Abbott Laboratories · Feb 19, 2010
Recalls
Recalls FDA ties to K980367, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record| May 1, 2019 | Z-1671-2020Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed. | Terminated |
| Aug 28, 2018 | Z-1296-2019There is a potential to generate falsely elevated serum or plasma chloride results when using the product. | Terminated |
| Jul 17, 2018 | Z-1104-2019, Z-1102-2019, Z-1103-2019There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status. | Terminated |
| Dec 8, 2017 | Z-0921-2018, Z-0922-2018, Z-0923-2018There is a potential to generate falsely depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000, c8000, and c16000 instrument. | Terminated |
| Sep 29, 2017 | Z-0166-2018, Z-0165-2018The Peristaltic Head tubing placed on the ARCHITECT c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector. | Completed |
| Sep 9, 2015 | Z-0163-2016Incorrect tubing. | Terminated |
| Jun 11, 2015 | Z-2202-2015The ARCHITECT c8000 instrument contains tubing that does not meet specifications and has the potential to leak. | Terminated |
| May 11, 2015 | Z-1725-2015The product is leaking from the cap and crystallization was noted as present around the cap. | Terminated |
| Jan 15, 2015 | Z-0011-2018There is a potential to generate falsely-depressed patient results in the cuvettes adjacent to the detached segment post for any assay run on the c4000 instrument. | Terminated |
| Jul 22, 2014 | Z-1932-2015ICT Serum Calibrator may generate lower than expected Potassium Quality Control (QC) and patient results on the Abbott Architect cSystem. | Terminated |
| May 23, 2008 | Z-2231-2008Software defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample. | Terminated |
| Sep 26, 2007 | Z-1166-2008Quality Control (Out of Range)- ICT Calibrator lot number: 0505017, may contain an interferent. Not all vials within a kit are affected. The investigation determined that the interferent affects Potassium (K+) only. Use… | Terminated |
| Sep 5, 2007 | Z-0146-2008Incorrect Sample/Test Identification: Due to a software timing defect with certain versions of ARCHITECT cSystem Software, test results may be incorrectly associated to a Sample Identification during specific… | Terminated |
| Apr 30, 2007 | Z-0952-2007One or more of four software issues : 1) calculation error for Architect cSystem assays using the Spline math model. 2) The system configuration option, Run controls onboard reagents by Kit, does not function correctly… | Terminated |
| Dec 8, 2005 | Z-0398-06System software assigns a calibrator default volume of 2.0uL when field is left empty by operator at time assay parameters set. Patient results could be affected if the volume required is not 2.0 uL. | Terminated |
| Sep 14, 2005 | Z-0010-06Software deficiency causing potential shift in absorbance readings for test results. | Terminated |
| Feb 28, 2005 | Z-0008-06Device has incorrect notation associated with therapeutic drug monitoring assays. | Terminated |
| Sep 24, 2004 | Z-0016-05Changes to the printer default settings made by users/operators cause printed data to be truncated and misidentified. | Terminated |
| Jun 10, 2004 | Z-1341-04Process change in the manufacturing of the AC/DC driver board used on the analyzer may result in one or more of the solid state relays to fail during use. | Terminated |
| Mar 22, 2004 | Z-0846-04Change of level 3 error log messages to level 1 to prevent suspect assay results. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 29, 1998
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1170
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files