Arca
Skip to content
Tools
Tools

Blood Gas, Electrolyte and Metabolite Test Card

Glucose Oxidase, Glucose. 510(k) K090109, from Epocal, Inc. (Ottawa, CA). Product code CGA, class 2.

As of Nov 18, 2026, K090109 is an FDA 510(k) held by Epocal, Inc., substantially equivalent on Jun 24, 2009 after 160 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 160 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K090109, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 13, 2013Z-1971-2013This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test…Terminated
Jun 28, 2011Z-2034-2012Customer reported problem while performing incoming QC of newly received lot of test cards. Liquid quality control results were found to be out of range. Further investigation by Epocal determined that a portion of the…Terminated
Oct 27, 2010Z-1287-2011Potential for out of specification star grind teeth.Terminated

Other CGA clearances

The newest 510(k)s for the same product code; their median review was 270 days; see every CGA device

Record

Type
Traditional
Received
Jan 15, 2009
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files