Carto 3 Ep Navigation System, V1.0 and Accessories

Computer, Diagnostic, Programmable. 510(k) K090017, from Biosense Webster, Inc. (Diamond Bar, CA). Product code DQK, class 2.

As of Nov 20, 2026, K090017 is an FDA 510(k) held by Biosense Webster, Inc., substantially equivalent on Oct 13, 2009 after 284 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 284 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The CARTO® 3 V1.0 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K090017, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Feb 9, 2012Z-1230-2012Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.Terminated
Feb 15, 2011Z-1839-2011The recall was initiated because Biosense Webster has confirmed that there is an inaccessibility of electronic version of the instruction for use (e-IFU) for the "10 ft. Interface Cable, 10 Pin-Shielded Tip Pins to 12…Terminated
Oct 5, 2010Z-0884-2011When connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.Terminated
Nov 17, 2009Z-2369-2012The recall was initiated because Biosense Webster has become aware of information regarding a potential hazard that may occur during the operation of a Bloom EP Stimulator device in conjunction with RF generators and…Terminated

Other DQK clearances

The newest 510(k)s for the same product code; their median review was 147 days; see every DQK device

Record

Type
Traditional
Received
Jan 2, 2009
Review panel
Cardiovascular
Regulation
21 CFR 870.1425

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files