Arca
Skip to content
Tools
Tools

Trilogy With External System Gating Interface

Accelerator, Linear, Medical. 510(k) K081188, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K081188 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Jul 14, 2008 after 77 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 77 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Trilogy TM Radiotherapy Delivery System is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. The External System Gating Interface is indicated for use in exporting beam information to external gating devices. 7.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K081188, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Aug 14, 2012Z-2383-2012Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the…Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Apr 28, 2008
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.