Trilogy Radiotherapy Delivery System and Trilogy TX Delivery System

Accelerator, Linear, Medical. 510(k) K070094, from Varian Medical Systems, Inc. (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K070094 is an FDA 510(k) held by Varian Medical Systems, Inc., substantially equivalent on Feb 9, 2007 after 30 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 30 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Trilogy $^{TM}$ Delivery System and Trilogy Tx Delivery System are indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070094, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 23, 2010Z-1249-2011Position errors may result in misdelivery of radiation.Terminated
Jun 10, 2010Z-0886-2011The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the…Terminated
Oct 8, 2009Z-0910-2011Product may deliver radiation treatment to areas larger than intended,to healthy tissue.Terminated
Sep 10, 2007Z-0544-2008Treatment Error: The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator…Terminated
Sep 10, 2007Z-0545-2008The collimator drive chain may break or slip off of its drive track, allowing the collimator to rotate freely without motor control; if undetected resulting in a treatment with the wrong collimator angle.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Special
Received
Jan 10, 2007
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files