Varian High Energy Linear Accelerator

Accelerator, Linear, Medical. 510(k) K100890, from Varian Medical Systems (Palo Alto, CA). Product code IYE, class 2.

As of Nov 19, 2026, K100890 is an FDA 510(k) held by Varian Medical Systems, substantially equivalent on Jul 13, 2010 after 104 days, with 11 recalls tied to it in FDA's recall database.

  • Review time: 104 days
  • Recalls: 11
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K100890, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 9, 2011Z-0181-2012Varian has received reports in which a user has remotely rotated the gantry into contact with the couch or with the patient, in both the manual mode and the automate mode.Terminated
May 31, 2011Z-1325-2013Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may…Terminated
May 26, 2011Z-2978-2011On Varian Linear Accelerators where the customer may calibrate the collimator angle position readout in reverse, switching the 90 degree and 270 degree positions.Terminated
May 24, 2011Z-2893-2011The throat cover on the High Energy Clinac may detach if not properly installed and possibly strike a patient.Terminated
May 18, 2011Z-2764-2011The Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.Terminated
May 13, 2011Z-2778-2011Some drive stand bolts may fail to meet mechanical specifications.Terminated
Mar 31, 2011Z-2224-2011The coolant may leak. The resulting coolant leak presents a risk of electrical shock to any person working within the protective housing.Terminated
Mar 28, 2011Z-2195-2011Varian has identified an anomaly whereby, following prolonged use, the screw fastener holding the wedge body to the tray may fail.Terminated
Mar 17, 2011Z-3204-2011An improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel.Terminated
Feb 24, 2011Z-2046-2011The wedge angle labeling on the wedge tray may not correctly match the wedge body. However, the wedge body is labeled with the correct wedge angle.Terminated
Dec 10, 2010Z-1045-2011An anomaly was identified whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Special
Received
Mar 31, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files