Accelerator, Linear, Medical. 510(k) K100890, from Varian Medical Systems (Palo Alto, CA). Product code IYE, class 2.
As of Nov 19, 2026, K100890 is an FDA 510(k) held by Varian Medical Systems, substantially equivalent on Jul 13, 2010 after 104 days, with 11 recalls tied to it in FDA's recall database.
- Review time: 104 days
- Recalls: 11
- Cited as a predicate by: 2 devices
The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Trilogy With External System Gating InterfaceK081188 · Varian Medical Systems, Inc. · Jul 14, 2008
- Trilogy System With RapidarcK072916 · Varian Medical Systems, Inc. · Nov 9, 2007
- Varian High Energy Linear AcceleratorK112839 · Varian Medical Systems · Nov 14, 2011
- Barcode Conical Collimator Verification (Bccv)K103394 · Varian Medical Systems, Oncology Systems · Jan 12, 2011
| Sep 9, 2011 | Z-0181-2012Varian has received reports in which a user has remotely rotated the gantry into contact with the couch or with the patient, in both the manual mode and the automate mode. | Terminated |
| May 31, 2011 | Z-1325-2013Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may… | Terminated |
| May 26, 2011 | Z-2978-2011On Varian Linear Accelerators where the customer may calibrate the collimator angle position readout in reverse, switching the 90 degree and 270 degree positions. | Terminated |
| May 24, 2011 | Z-2893-2011The throat cover on the High Energy Clinac may detach if not properly installed and possibly strike a patient. | Terminated |
| May 18, 2011 | Z-2764-2011The Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing. | Terminated |
| May 13, 2011 | Z-2778-2011Some drive stand bolts may fail to meet mechanical specifications. | Terminated |
| Mar 31, 2011 | Z-2224-2011The coolant may leak. The resulting coolant leak presents a risk of electrical shock to any person working within the protective housing. | Terminated |
| Mar 28, 2011 | Z-2195-2011Varian has identified an anomaly whereby, following prolonged use, the screw fastener holding the wedge body to the tray may fail. | Terminated |
| Mar 17, 2011 | Z-3204-2011An improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel. | Terminated |
| Feb 24, 2011 | Z-2046-2011The wedge angle labeling on the wedge tray may not correctly match the wedge body. However, the wedge body is labeled with the correct wedge angle. | Terminated |
| Dec 10, 2010 | Z-1045-2011An anomaly was identified whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Mar 31, 2010
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files