Neonatal Bilirubin On Rapidlab Models 1245 and 1265
Bilirubin (Total And Unbound) In The Neonate Test System. 510(k) K073537, from Siemens Medical Solutions Diagnostics (Norowood, MA). Product code MQM, class 1.
As of Nov 18, 2026, K073537 is an FDA 510(k) held by Siemens Medical Solutions Diagnostics, substantially equivalent on Jun 27, 2008 after 193 days, with 7 recalls tied to it in FDA's recall database.
- Review time: 193 days
- Recalls: 7
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe neonatal bilirubin test intended use on the Rapidlab 1245 and Rapidlab 1265 analyzers is an in vitro diagnostic test for the determination of total neonatal bilirubin (nBili) concentration in the whole blood of newborn infants. Measurement of nBili aids in assessing the risk of kernicterus.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- ABL800 Flex With Flexq Module, Model ABL8XXK043218 Ā· Radiometer Medical Aps Ā· May 10, 2005
- Abl 735 AnalyzerK991417 Ā· Radiometer America, Inc. Ā· Nov 24, 1999
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Rapidpoint 405 Neonatal BilirubinK110277 Ā· Siemens Healthcare Diagnostics Ā· Aug 15, 2011
Recalls
Recalls FDA ties to K073537, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Aug 11, 2016 | Z-2803-2016, Z-2802-2016There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system | Terminated |
| May 6, 2015 | Z-1962-2015D50 and D51 diagnostic error codes are not functional. | Terminated |
| Dec 30, 2014 | Z-1255-2015, Z-1254-2015Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl. | Terminated |
| Sep 17, 2012 | Z-0107-2014, Z-0108-2014RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the⦠| Terminated |
Record
- Document
- 510(k) summary (PDF) ā
- Type
- Traditional
- Received
- Dec 17, 2007
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.1113
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files