Arca
Skip to content
Tools
Tools

Rapidpoint 405 Neonatal Bilirubin

Bilirubin (Total And Unbound) In The Neonate Test System. 510(k) K110277, from Siemens Healthcare Diagnostics (Norowood, MA). Product code MQM, class 1.

As of Nov 18, 2026, K110277 is an FDA 510(k) held by Siemens Healthcare Diagnostics, substantially equivalent on Aug 15, 2011 after 196 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 196 days
  • Recalls: 5

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K110277, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Sep 2, 2016Z-2913-2016Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement CartridgeTerminated
Aug 11, 2016Z-2798-2016There is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the systemTerminated
Jul 28, 2015Z-2479-2015When both ports (Serial and Ethernet) are configured to transmit data, the data stream from one port could potentially affect the data stream from the other port causing the data management system to include either:…Terminated
Dec 30, 2014Z-1252-2015Neonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.Terminated
Sep 17, 2012Z-0103-2014RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the…Terminated

Record

Type
Traditional
Received
Jan 31, 2011
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1113

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.