Evis Exera 180 System

Bronchoscope (Flexible Or Rigid). 510(k) K061313, from Olympus Medical Systems Corporation (Center Valley, PA). Product code EOQ, class 2.

As of Nov 17, 2026, K061313 is an FDA 510(k) held by Olympus Medical Systems Corporation, substantially equivalent on Aug 31, 2006 after 113 days, with 19 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Aug 31, 2006
  • Review time: 113 days
  • Recalls: 19
  • Cited as a predicate by: 7 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K061313, newest first; each opens FDA's recall record
Jan 29, 2026Z-1489-2026Potential for rubber fragment detachment during use.Open, Classified
Jan 29, 2026Z-1488-2026Potential for rubber fragment detachment during use.Open, Classified
Sep 11, 2025Z-0046-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0051-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0056-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Sep 11, 2025Z-0061-2026Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.Open, Classified
Oct 12, 2023Z-0191-2024There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the…Open, Classified
Sep 25, 2023Z-0332-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified
Jul 27, 2023Z-2493-2023Complaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.Open, Classified
Jun 8, 2023Z-2016-2023There have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.Open, Classified
Mar 8, 2022Z-1049-2022Bronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using…Open, Classified
Aug 16, 2021Z-2571-2021The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."Terminated
Aug 16, 2021Z-2572-2021The firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."Terminated
Aug 31, 2020Z-3017-2020The BF-Q180 bronchoscope is being recalled because it does not have a 510(k) clearance, and the device is associated with a higher rate of patient infections than other comparable OMSC bronchoscopes.Terminated
Aug 8, 2018Z-3288-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated
Aug 8, 2018Z-3301-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated
Aug 8, 2018Z-3297-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated
Aug 8, 2018Z-3285-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated
Aug 8, 2018Z-3310-2018The attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.Terminated

Other EOQ Clearances

The newest 510(k)s for the same product code; their median review was 229 days; see every EOQ device

Record

Type
Traditional
Received
May 10, 2006
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4680

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files