Rhino-Laryngo Videoscope Olympus ENF-VT3

Nasopharyngoscope (Flexible Or Rigid). 510(k) K183416, from Olympus Medical Systems Corp. (Hachioji-Shi, JP). Product code EOB, class 2.

As of Nov 17, 2026, K183416 is an FDA 510(k) held by Olympus Medical Systems Corp., substantially equivalent on Sep 6, 2019 after 270 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: Sep 6, 2019
  • Review time: 270 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

This instrument is intended to be used with an Olympus video system center, light somce, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment. This instrument is indicated for use within the nasal lumens and airway anatomy (including nasopharynx and trachea).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K183416, newest first; each opens FDA's recall record
StartedReasonStatus
Sep 25, 2023Z-0332-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for…Open, Classified

Other EOB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every EOB device

Record

Type
Traditional
Received
Dec 10, 2018
Review panel
Ear, Nose, Throat
Regulation
21 CFR 874.4760

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files