Evis Exeraii 180 System
Endoscope, Accessories, Narrow Band Spectrum. 510(k) K100584, from Olympus Medical Systems Corporation (Center Valley, PA). Product code NWB, class 2.
As of Nov 17, 2026, K100584 is an FDA 510(k) held by Olympus Medical Systems Corporation, substantially equivalent on Jul 2, 2010 after 122 days, with 9 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 2, 2010
- Review time: 122 days
- Recalls: 9
- Cited as a predicate by: 19 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFEVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Evis Exera 180 SystemK061313 · Olympus Medical Systems Corporation · Aug 31, 2006
- Evis Exera Ii 180 SystemK062049 · Olympus Medical Systems Corporation · Sep 22, 2006
- Evis Exera 160A SystemK051645 · Olympus Medical Systems Corporation · Oct 13, 2005
Cited as a Predicate
19 later 510(k) summaries name this device as a predicate; the 12 newest- Triton 1 SystemK260503 · Neptune Medical, Inc. · Sep 3, 2026
- Endoscopic Video Image Processor (RP-IPD-V2000A, RP-IPD-V2000B, RP-IPD-V2000C,…K232043 · Guangzhou Red Pine Medical Instrument Co., Ltd. · Feb 16, 2024
- Aer-O-Scope Colonoscope SystemK230588 · Gi View , Ltd. · Aug 17, 2023
- Ambu® aScope™ Colon; Ambu® aBox™ 2K230332 · Ambu A/S · Sep 15, 2023
- Ambu® aScope™ Gastro, Ambu® aBox™2K212382 · Ambu A/S · Feb 3, 2022
- Saneso Colonoscope 360-A (Model: with/without Select Band Imaging (SBI) and with/without…K210052 · Saneso, Inc. · Apr 1, 2022
- Ambu Duodeno SystemK201098 · Ambu Innovation GmbH · Jul 17, 2020
- Endofresh Digestive Endoscopy SystemK192704 · Huizhou Xzing Technology Co., Ltd. · Sep 2, 2020
- HD-500 Video Endoscope SystemK173921 · Sonoscape Medical Corp. · Sep 5, 2018
- invendoscopy E210 SystemK173085 · Invendo Medical GmbH · Jan 8, 2018
- invendoscopy E200 SystemK161355 · Invendo Medical GmbH · Aug 25, 2016
- EPX-4440HD Digital Video ProcessorK140149 · Fujifilm Medical Systems U.S.A, Inc. · Aug 22, 2014
Recalls
Recalls FDA ties to K100584, newest first; each opens FDA's recall record| Jan 11, 2024 | Z-1209-2024Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads. | Open, Classified |
| Jan 11, 2024 | Z-1205-2024Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads. | Open, Classified |
| Jan 11, 2024 | Z-1208-2024Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads. | Open, Classified |
| Jan 11, 2024 | Z-1214-2024Specific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in temperature or severe shock loads. | Open, Classified |
| Sep 25, 2023 | Z-0334-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Sep 25, 2023 | Z-0329-2024Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for… | Open, Classified |
| Dec 21, 2022 | Z-1132-2023A single CF-Q180AL colonovideoscope was utilized in a veterinary endoscopy procedure in advance of being assigned to a medical facility as a service loaner in error, potential for microbial contamination | Open, Classified |
| Apr 22, 2022 | Z-1274-2022A (Camera unit) CCD intended for a different model of colonoscope may have been installed in the CF-H180AL, viewing angle would be narrower than a normal CF-H180AL and the use of EndoTherapy devices may stick out… | Terminated |
| Dec 17, 2018 | Z-1009-2019The adhesive used in the repair of the endoscope was incorrectly prepared. It is unknown what affect the nonconforming adhesive mixture would have on the durability or effectiveness of the gluing operation for the… | Terminated |
Other NWB Clearances
The newest 510(k)s for the same product code; see every NWB device- HD-550 Video Endoscope SystemK222020 · Sonoscape Medical Corp. · Sep 8, 2022
- HD-550 Video Endoscope SystemK211882 · Sonoscape Medical Corp. · Apr 1, 2022
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 2, 2010
- Review panel
- Gastroenterology, Urology
- Regulation
- 21 CFR 876.1500
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files