Arca
Skip to content
Tools
Tools

Nexgen Knee Gender Solutions Female (Gsf) Femoral Components

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K060370, from Zimmer, Inc. (Warsaw, IN). Product code JWH, class 2.

As of Nov 19, 2026, K060370 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Apr 28, 2006 after 74 days, with 17 recalls tied to it in FDA's recall database.

  • Review time: 74 days
  • Recalls: 17
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K060370, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 29, 2017Z-1056-2018, Z-1058-2018, Z-1059-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Jan 11, 2016Z-0830-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
Aug 30, 2013Z-0073-2014, Z-0074-2014The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.Terminated
Sep 15, 2010Z-0535-2011, Z-0532-2011, Z-0530-2011, Z-0536-2011, Z-0528-2011, Z-0531-2011, Z-0533-2011, Z-0529-2011, Z-0534-2011These lots of Zimmer NexGen Complete Knee Solution LPS Femoral Component and Zimmer NexGen Complete Knee Solution LPS Flex Gender Femoral Component exhibit a nonconforming internal CAM radius. The condition is…Terminated
Jan 27, 2010Z-0915-2010The package should contain the left knee component but actually contains the right knee component. Delay in completing surgery may result, and a new or different implant may need to be identified.Terminated
Nov 13, 2009Z-0501-2010The inner package may interfere with the outer seal, resulting in lack of assurance of sterility.Terminated

Other JWH clearances

The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device

Record

Type
Traditional
Received
Feb 13, 2006
Review panel
Orthopedic
Regulation
21 CFR 888.3560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files