Arca
Skip to content
Tools
Tools

Nexgen Complete Knee Solution Lps-Flex and Knee Gender Solutions Female (Gsf) Porous Femoral Components

Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer. 510(k) K072619, from Zimmer, Inc. (Warsaw, IN). Product code MBH, class 2.

As of Nov 19, 2026, K072619 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Nov 21, 2007 after 65 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 65 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

General Indications for the NexGen Complete Knee Solution Femoral Components: - This device is indicated for patients with severe knee pain and disability due to: - This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K072619, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 29, 2017Z-1044-2018, Z-1057-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Jan 11, 2016Z-0840-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
Apr 16, 2010Z-0609-2011The impactor may disassemble or detach during use. The detached piece may not be observed intraoperatively and may be left in the surgical site. Complications may include pain, infection, and prolongation in surgery.…Terminated

Other MBH clearances

The newest 510(k)s for the same product code; their median review was 153 days; see every MBH device

Record

Type
Traditional
Received
Sep 17, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3565

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files