Zimmer Patellofemoral Joint Prosthesis
Prosthesis, Knee, Patello/Femoral, Semi-Constrained, Cemented, Metal/Polymer. 510(k) K070695, from Zimmer, Inc. (Warsaw, IN). Product code KRR, class 2.
As of Nov 18, 2026, K070695 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Jun 7, 2007 after 86 days, with 5 recalls tied to it in FDA's recall database.
- Review time: 86 days
- Recalls: 5
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFOsteoarthritis, traumatic arthritis, polyarthritis, and/or severe chondrocalcinosis of the patellofemoral joint; - The salvage of previously failed surgical attempts (e.g., arthroscopy, lateral release, cartilage transplantation), -History of patellar dislocation or patella fracture; -Dysplasia-induced degeneration This device is intended for cemented use only.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Stryker Compartmental Knee SystemK052917 · Howmedica Osteonics Corp. · Dec 27, 2005
- Natural-Knee Ii Patellofemoral Joint ProsthesisK002356 · Sulzer Orthopedics, Inc. · Oct 30, 2000
- Nexgen Knee Gender Solutions Female (Gsf) Femoral ComponentsK060370 · Zimmer, Inc. · Apr 28, 2006
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- MOTO PFJ SystemK200122 · Medacta International S.A. · Apr 20, 2020
- Gender Solutions Natural-Knee Flex SystemK073286 · Zimmer, Inc. · Mar 7, 2008
Recalls
Recalls FDA ties to K070695, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Nov 29, 2017 | Z-1053-2018, Z-1052-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. | Terminated |
| Jan 11, 2016 | Z-0837-2016LDPE bag containing the implant adheres to the highly polished implant surface. | Terminated |
| Jul 30, 2014 | Z-2545-2014Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to… | Terminated |
| Aug 30, 2013 | Z-0077-2014The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices. | Terminated |
Other KRR clearances
The newest 510(k)s for the same product code; their median review was 67 days; see every KRR device- Overture Orthopaedics Patellofemoral SystemK231253 · Overture Resurfacing, Inc. · Jul 7, 2023
- Episealer Patellofemoral SystemK221048 · Episurf Medical, Inc. · Dec 23, 2022
- Avon Patello-femoral Joint ProsthesisK211303 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jun 4, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Mar 13, 2007
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3540
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files