Prosthesis, Knee, Patello/Femoral, Semi-Constrained, Cemented, Metal/Polymer. 510(k) K070695, from Zimmer, Inc. (Warsaw, IN). Product code KRR, class 2.

As of Nov 18, 2026, K070695 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Jun 7, 2007 after 86 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 86 days
  • Recalls: 5
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

Osteoarthritis, traumatic arthritis, polyarthritis, and/or severe chondrocalcinosis of the patellofemoral joint; - The salvage of previously failed surgical attempts (e.g., arthroscopy, lateral release, cartilage transplantation), -History of patellar dislocation or patella fracture; -Dysplasia-induced degeneration This device is intended for cemented use only.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K070695, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 29, 2017Z-1053-2018, Z-1052-2018The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.Terminated
Jan 11, 2016Z-0837-2016LDPE bag containing the implant adheres to the highly polished implant surface.Terminated
Jul 30, 2014Z-2545-2014Internal review found that five (5) PFJ implant forgings were identified as have forged laps during receiving inspection process for raw material. These five forgings were not segregated and were subsequently issued to…Terminated
Aug 30, 2013Z-0077-2014The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.Terminated

Other KRR clearances

The newest 510(k)s for the same product code; their median review was 67 days; see every KRR device

Record

Type
Traditional
Received
Mar 13, 2007
Review panel
Orthopedic
Regulation
21 CFR 888.3540

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files