Spacelabs Medical Multigas Analyzer Module 91518 And Accessories
Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (Anesthetic Conc.). 510(k) K053599, from Spacelabs Medical, Inc. (Issaquah, WA). Product code CBR, class 2.
As of Nov 18, 2026, K053599 is an FDA 510(k) held by Spacelabs Medical, Inc., substantially equivalent on Apr 19, 2006 after 117 days, with 2 recalls tied to it in FDA's recall database.
- Substantially equivalent: Apr 19, 2006
- Review time: 117 days
- Recalls: 2
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe Spacelabs Medical Multigas Analyzer Module 91518 (Module 91518) is intended to provide a means of monitoring a variety of gas concentrations and to alert clinical personnel when the concentration of anesthetic agent, oxygen, carbon dioxide or nitrous oxide falls outside of user defined limits.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Spacelabs Medical Multigas AnalyzerK954962 · Spacelabs, Inc. · May 31, 1996
- Datex-Ohmeda S/5 Compact Airway Module (Model Family E-Caiovx) E-Caiovx, E-Caiov, E-Caio, E-Covx, E-Cov, E-Co & AccesaccK051092 · GE Healthcare · May 25, 2005
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Speacelabs Multigas ModuleK112173 · Spacelabs Healthcare · Oct 5, 2011
Recalls
Recalls FDA ties to K053599, newest first; each opens FDA's recall record| Jun 8, 2011 | Z-2902-2011Spacelabs received a report of the Model 91518 Multigas Analyzer trending low levels of an anesthetic agent that is not present during a particular case. The error is limited to only the trend table data. | Terminated |
| Oct 8, 2007 | Z-0660-2008Inaccurate partial pressure readings; readings are not compensated for ambient pressure in mmHg or kPa. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Dec 23, 2005
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.1700
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files