Arca
Skip to content

Datex-Ohmeda S/5 Compact Airway Module (Model Family E-Caiovx) E-Caiovx, E-Caiov, E-Caio, E-Covx, E-Cov, E-Co & Accesacc

Analyzer, Gas, Oxygen, Gaseous-Phase. 510(k) K051092, from GE Healthcare (Needham, MA). Product code CCL, class 2.

As of Nov 17, 2026, K051092 is an FDA 510(k) held by GE Healthcare, substantially equivalent on May 25, 2005 after 27 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: May 25, 2005
  • Review time: 27 days
  • Recalls: 2
  • Cited as a predicate by: 8 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Datex-Ohmeda S/5™ Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patient's respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas exchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE).

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

8 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K051092, newest first; each opens FDA's recall record
Apr 24, 2015Z-1680-2015A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fend / D-Fend+ water traps used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. This water trap…Terminated
May 23, 2014Z-1808-2014Potential loss of airway gas measurement in the Compact Airway Gas Modules. Loosening of the gas sampling pump hex screw in the module pump unit may lead to pump failure that may cause loss of respiratory airway gas…Terminated

Other CCL Clearances

The newest 510(k)s for the same product code; their median review was 112 days; see every CCL device

Record

Type
Special
Received
Apr 28, 2005
Review panel
Anesthesiology
Regulation
21 CFR 868.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.