Arca
Skip to content
Tools
Tools

Speacelabs Multigas Module

Analyzer, Gas, Nitrous-Oxide, Gaseous Phase (Anesthetic Conc.). 510(k) K112173, from Spacelabs Healthcare (Nederland, CO). Product code CBR, class 2.

As of Nov 19, 2026, K112173 is an FDA 510(k) held by Spacelabs Healthcare, substantially equivalent on Oct 5, 2011 after 69 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 69 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Spacelabs Multigas Module, Model 92518 (Model 92518) is intended to provide a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits. The multigas module is capable of automatically identifying which anesthetic agent(s) is being administered.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K112173, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 18, 2014Z-2077-2014With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.Terminated

Record

Type
Special
Received
Jul 28, 2011
Review panel
Anesthesiology
Regulation
21 CFR 868.1700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files