Mac-Lab/Cardiolab Ep/Combolab System
Computer, Diagnostic, Programmable. 510(k) K050093, from Ge Medical Systems Information Technologies (Milwaukee, WI). Product code DQK, class 2.
As of Nov 20, 2026, K050093 is an FDA 510(k) held by Ge Medical Systems Information Technologies, substantially equivalent on May 13, 2005 after 119 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 119 days
- Recalls: 4
- Cited as a predicate by: 2 devices
Mac-Lab System: The Mac-Lab System is intended for use in a catheterization and related cardiovascular specialty laboratories under the direct supervision of a licensed healthcare practitioner. It is intended to monitor, calculate and/or record cardiovascular data from adult and pediatric patients undergoing cardiac catheterization procedures.
- Modification To: Mac-Lab/Cardiolab Ep/Combolab SystemK032577 · Ge Medical Systems Information Technologies · Oct 10, 2003
- Xper Information Management and Flex Cardio Physiomonitoring SystemsK101571 · Witt Biomedical Corp · Oct 26, 2010
- Mac-Lab/Cardiolab /Combolab/Specialslab SystemK061741 · GE Healthcare · Sep 15, 2006
| Apr 22, 2024 | Z-2258-2024Capacitors in certain Prucka 3 Amplifiers used with CardioLab / ComboLab systems could fail resulting in transient oscillations on the display and inability to view surface and intracardiac ECG waveforms. | Open |
| Feb 16, 2007 | Z-1133-2007, Z-1132-2007, Z-1134-2007GE Healthcare has identified conditions that could cause a potential loss of real-time monitoring with the CardioLab/Mac-Lab/ComboLab Electrophysiology and Hemodynamic Monitoring Systems. These issues are associated… | Terminated |
- Merge Hemo (Model RCSV2 (v11.3))K262171 · Merge Healthcare Incorporated · Nov 5, 2026
- Vivo (9002)K261886 · Catheter Precision, Inc. · Oct 30, 2026
- PulseAPI SmartWatchK254226 · PulseAI, Ltd. · Oct 27, 2026
- STAR Apollo™ Mapping SystemK263380 · Rhythm AI, Ltd. · Oct 23, 2026
- Intra-Operative Positioning System (IOPS) and IOPS Visionary SystemK263016 · Centerline Biomedical · Oct 1, 2026
- S-Patch CardioAIK254255 · Wellysis Corp. · Sep 25, 2026
- EnSite™ X EP SystemK261729 · Abbott Medical · Sep 17, 2026
- AVVIGO™+ Multi-Modality Guidance SystemK253692 · Boston Scientific · Aug 21, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jan 14, 2005
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1425
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files