Unicel Dxi 800 Access Immunoassay System, Model 973100
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use. 510(k) K023764, from Beckman Coulter, Inc. (Chaska, MN). Product code JJE, class 1.
As of Nov 18, 2026, K023764 is an FDA 510(k) held by Beckman Coulter, Inc., substantially equivalent on Jan 28, 2003 after 77 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 77 days
- Recalls: 25+
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe UniCel Dxl 800 Access Immunoassay System is a microcomputer controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Access Theophylline Immunoassay ReagentsK922823 · Bio-Rad Laboratories, Inc. · Aug 5, 1992
- Access Afp Immunoassay SystemP980041 · Beckman Coulter, Inc. · Feb 8, 1999
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- DxA 5000, DxI 800 Access Immunoassay System, Access Ferritin, Access Folate, Access TSH (3rd IS), Access Vitamin B12K190298 · Beckman Coulter Biomedical GmbH · Oct 4, 2019
- Access Accutni Reagent, Access Accutni Calibrator, Unicel Dxi 800 Access Immunoassay SystemK121790 · Beckman Coulter, Inc. · Sep 25, 2013
- Power Processor Sample Processing System With Generic Connection Module, Model: 4210K090588 · Beckman Coulter, Inc. · May 27, 2009
Recalls
Recalls FDA ties to K023764, newest first; the 25 newest; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jun 25, 2025 | Z-2271-2025, Z-2268-2025, Z-2270-2025, Z-2266-2025, Z-2269-2025, Z-2267-2025Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current… | Open |
| Jun 20, 2023 | Z-2237-2023Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an… | Open |
| Dec 5, 2018 | Z-1077-2019When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an… | Open |
| Sep 18, 2018 | Z-2763-2019Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%). | Terminated |
| Aug 21, 2014 | Z-2687-2014Beckman Coulter is recalling the Sample Probe Cable because nonconforming sample probe cables may been installed on newly manufactured and/or recently serviced UniCel DxI 600 and 800 Access Immunoassay Systems, UniCel… | Terminated |
| Nov 26, 2013 | Z-0985-2014The Access 2 Immunoassay System Analyzer and the UniCel DxI 600 Access may have been manufactured with non-conforming bushings. Non-conforming bushings have a remote probability of generating incorrect results. | Terminated |
| Oct 13, 2011 | Z-2259-2012, Z-2257-2012, Z-2258-2012The recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results. | Terminated |
| Aug 1, 2011 | Z-0346-2013, Z-0345-2013The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies. | Terminated |
| Apr 15, 2011 | Z-3127-2011The recall was initiated because certain Access immunoassays can produce inaccurate results when performed within the published UniCel DxI Immunoassay Systems* room temperature operating specification of 18 degrees C to… | Terminated |
| Jan 26, 2011 | Z-1938-2011, Z-1940-2011, Z-1939-2011The recall was initiated due to Beckman Coulter receiving a report of results that were associated with incorrect patient identifications when two sample racks advanced into the sample presentation area in one push. The… | Terminated |
| Aug 13, 2010 | Z-0607-2011This recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated… | Terminated |
| Jul 6, 2010 | Z-2685-2011Test results may have incorrect ranges. If a default sample type setting on the tests screen is changed while running Unicel DXI system software version 4.3, the new setting is not saved to the system database. The… | Terminated |
| Oct 28, 2009 | Z-0856-2010This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red… | Terminated |
| Oct 9, 2008 | Z-0914-2010The recall was initiated after Beckman Coulter confirmed reports of a new premature failure of the upper aspirate peristaltic pump in the systems listed above. Premature failure of the pump may negatively affect… | Terminated |
| Apr 3, 2008 | Z-2387-2008The field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively… | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Nov 12, 2002
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.2160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files