Access Afp Immunoassay System
Kit, test, alpha-fetoprotein for neural tube defects. PMA P980041, from Beckman Coulter, Inc. (Brea, CA). Product code LOK, class 3.
As of Nov 18, 2026, P980041 is an FDA PMA held by Beckman Coulter, Inc., approved on Feb 8, 1999 after 115 days, with no recalls tied to it in FDA's recall database.
- Review time: 115 days
- Supplements: 47
- Recalls: 0
- Cited as a predicate by: 1 device
What FDA approved
FDA's approval order statementApproval for the Access(R) AFP Reagents on the Access(R) Immunoassay Analyzer. The Access(R) AFP assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access(R) Immunoassay System for the quantitative determination fo alpha-fetoprotein (AFP) in: 1)Human seru, as an aid in management of patients with non-seminomatous testicular cancer. 2) Maternal serum and amniotic fluid at 15 to 20 weeks gestation, to aid in the detection of fetal open neural tube defects (ONTD). The assay is intended for use in conjunction with other diagnostic tools such as ultrasound and amniography.
Pivotal study
The main clinical evidence FDA relied on, as its Summary of Safety and Effectiveness describes it- Design
- Multicenter, retrospective study of archived maternal serum and amniotic fluid specimens vs. confirmed pregnancy outcomes
- Enrolled
- 3,737
- Where
- 3 sites, US, Canada
- Primary endpoint
- Clinical sensitivity and specificity for detection of fetal open neural tube defects (ONTD) at 15-20 weeks gestation
- Result
- Maternal serum at 2.0 MoM: sensitivity 91.3% (21/23), specificity 95.4% (2421/2539); amniotic fluid at 3.0 MoM: sensitivity 100% (15/15), specificity 99.3% (715/720)
Cited as a predicate
1 later 510(k) summary names this device as a predicate- Unicel Dxi 800 Access Immunoassay System, Model 973100K023764 · Beckman Coulter, Inc. · Jan 28, 2003
Supplements
47 supplements to this PMA; the 15 newest| Nov 3, 2025 | S051Labeling Change - Indications/instructions/shelf life/tradename | Normal 180 Day Track |
| Dec 16, 2022 | S050Change Design/Components/Specifications/Material | Real-Time Process |
| Dec 16, 2020 | S049Change Design/Components/Specifications/Material | Real-Time Process |
| Oct 16, 2020 | S048Change Design/Components/Specifications/Material | Real-Time Process |
| Oct 7, 2020 | S047Change Design/Components/Specifications/Material | Real-Time Process |
| May 28, 2020 | S043Process Change - Manufacturer/Sterilizer/Packager/Supplier | 135 Review Track For 30-Day Notice |
| Mar 17, 2020 | S046Change Design/Components/Specifications/Material | Real-Time Process |
| Aug 9, 2019 | S045Change Design/Components/Specifications/Material | Real-Time Process |
| Jul 31, 2019 | S044Change Design/Components/Specifications/Material | Real-Time Process |
| Dec 21, 2018 | S041Change Design/Components/Specifications/Material | Real-Time Process |
| Sep 12, 2018 | S040Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Oct 24, 2017 | S039Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Jul 7, 2017 | S037Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
| Jun 16, 2017 | S036Change Design/Components/Specifications/Material | Real-Time Process |
| Feb 14, 2017 | S035Process Change - Manufacturer/Sterilizer/Packager/Supplier | 30-Day Notice |
Record
- Document
- Approval order (PDF) ↗
- Type
- Original PMA
- Received
- Oct 16, 1998
- Review panel
- Immunology
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files