WARNING LETTER CMS # 732874
July 16, 2026
Dear Mr. Yamazaki:
During an inspection of your firm located in Lewisberry, PA from February 2, 2026 through March 27, 2026, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures dry and liquid hemodialysis dialysate concentrate solutions. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or any function of the body.
Quality Management System Regulation Violations This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality Management System Regulation found at Title 21, Code of Federal Regulations (CFR), Part 820.
We received responses from Rachael Nelsen, Vice President of Quality Affairs & Regulatory Affairs, dated April 17, 2026 and May 29, 2026, and from Vikram Chandra, Senior Vice President of Quality Affairs & Regulatory Affairs, dated July 14, 2026 concerning our investigator’s observations noted on the Form FDA 483 (FDA 483), List of Inspectional Observations, that was issued to your firm. We address these responses below, in relation to each of the noted violations. These violations include, but are not limited to, the following:
1. Failure to ensure that product which does not conform to product requirements is identified and controlled to prevent its unintended use or delivery; and failure to document a procedure defining the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product, as required by ISO 13485:2016, Clause 8.3.1. Specifically,
a. Your firm released and distributed finished hemodialysis products with out-of-specification (OOS) endotoxin results on multiple occasions. Specifically, your firm released the following lots of products despite your Laboratory Report Forms (DMR 5404) incorrectly indicated that endotoxin LAL testing had 'Passed,' and the Finished Product Packet (SOP QSM012; Effective Date 04/28/23) contained Quality's signed acceptance of each lot, despite each example exceeding the (b)(4) /mL endotoxin specification:
• On May 6, 2025, MedicaLyte Bicarbonate Powder (45X), Lot (b)(4) (expiration 04/30/2027), was released with an endotoxin result of (b)(4) /mL
• On December 18, 2025, MedicaLyte Bicarbonate Powder (DB+201-25), Lot (b)(4) (expiration 12/05/2027), was released with an endotoxin result of (b)(4) /mL
• On May 22, 2025, Citric Complete Dry Citric Acid Concentrate, Lot (b)(4) (expiration 05/19/2027), was released with an endotoxin result of (b)(4) /mL
b. Your firm's Investigations procedure (NRSU-SOP-0060, Revision 02, Effective Date December 8, 2025) and the Endotoxin Analysis work instruction (NRSU-WI-0061, Revision 02, Effective Date October 31, 2025, and Revision 01, Effective Date August 25, 2025) define processes for investigating OOS results. Section 7.2 of NRSU-SOP-0060 requires a MasterControl form for each investigation initiated for OOS results and issues found during production. However, an OOS log for January 2025 through February 2026, generated at investigators' request during the inspection, revealed thirteen product-related endotoxin failures and (b)(4) endotoxin failures for which your firm was unable to provide any documented investigation or nonconformance documentation.
We reviewed your firm's responses and conclude that they are not adequate. Your firm provided Health Hazard Assessments (NRSU-RPT-0028 Rev 02 and NRSU-RPT-0045) asserting that retain samples of the above lots yielded compliant endotoxin results, and therefore no field action or customer notification is warranted, attributing the OOS documentation to documentation errors. However, retesting of retain samples is not an appropriate justification for not performing a field action or customer notification when the original test results demonstrated product with endotoxin test results exceeding specification. Furthermore, our Agency has reviewed these results and your firm’s response (including the Health Hazard Assessment and Customer Notification Determination) and has determined that these lots of products with endotoxin levels higher than specification may cause a risk to health, namely (b)(4) reactions. In your response, please provide what your firm’s planned action(s) are related to these adulterated, released lots of products.
Additionally, your firm initiated CAPAs 26004 and 26005, with promised corrective actions including revisions to NRSU-SOP-0060 to establish clear, enforceable, and risk-based criteria for investigation initiation and escalation across all applicable quality events — including laboratory OOS results, (b)(4) and utility monitoring failures, and nonconforming product. However, these items remain in progress as of your July 14, 2026 response. In response to this letter, provide documentation for any outstanding corrective actions, including verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
2. Failure to take action to eliminate the cause of nonconformities in order to prevent recurrence, failure to take any necessary corrective actions without undue delay, and failure to take corrective actions proportionate to the effects of the nonconformities encountered, as required by ISO 13485:2016, Clause 8.5.2.
Specifically, your firm's Corrective and Preventive Action (CAPA) procedure (NRSU-SOP-0059, Revision 01, Effective Date 10/08/2024) has not been fully implemented in that corrective actions have not been proportionate to the effects of the nonconformities encountered, have not been taken without undue delay, and have not prevented recurrence of nonconformities. For example:
• CAPA-25012 (initiated June 13, 2025, for endotoxin documentation failures) was found to be ineffective when your firm discovered that Lot (b)(4), MedicaLyte Bicarbonate Powder (DB+201-25), was released with an OOS endotoxin result after corrective actions had been implemented, demonstrating recurrence of the nonconformance.
• CAPA-25014 (initiated August 4, 2025 to address facility and equipment maintenance deficiencies including corrosion, contamination, and poor environmental controls in the (b)(4) room) was extended on January 19, 2026, to a revised due date of March 30, 2026. Despite this, during a facility tour on February 13, 2026 ((b)(4) days after CAPA initiation), product residue was observed to be accumulated on and suspended from bottle filling machine nozzles, demonstrating that necessary corrective actions had not been taken without undue delay.
• CAPA-25005 (initiated May 22, 2025, for failing to follow the Monitoring Production Systems procedure) had a corrective action expected completion date of April 9, 2026 which is approximately (b)(4) days from initiation. Despite this, review of quality packets from September 2025, November 2025, and January 2026 during the inspection found that maintenance, sanitization, and (b)(4) monitoring forms continued to be missing and/or incomplete, demonstrating that necessary corrective actions had not been taken without undue delay.
• CAPA-25010 (initiated June 4, 2025, for failure to follow and perform sampling of the (b)(4) system per NRSU-SOP-0039) corrective actions were completed on January 6, 2026. However, inspection review of January 2026 (b)(4) monitoring records revealed that multiple sample port tests and (b)(4) had not been conducted for colony count and endotoxin monitoring without any documented justification, demonstrating that your firm failed to eliminate the cause of nonconformities to prevent recurrence.
The adequacy of your firm's responses cannot be determined at this time. Your firm stated that CAPAs 25005, 25010, 25014, and 25015 had all been initiated prior to the current FDA inspection. Your firm's response promises revisions to NRSU-SOP-0059, implementation of a CAPA tracking system, and training for applicable personnel, with a target completion date of August 7, 2026. CAPA-25015, which directly addresses your CAPA process itself, remains open. None of the underlying nonconformances addressed by the above CAPAs have been resolved, and all remain ongoing or have reoccurred. In response to this letter provide documentation for any status updates for the outstanding corrective actions, including verification of effectiveness, when completed. If corrective actions are not complete, provide a timeframe for completion.
3. Failure to document procedures for design and development, as required by ISO 13485:2016, Clause 7.3.1.
Specifically, although your firm maintains procedure NRSU-SOP-0018, Design Control (Revision 00, Effective Date 05/01/2020), and your Quality Manual (NRSU-MAN-0001, Revision 01, Effective Date 04/21/2025) describe design and development requirements including design and development planning, design input, design output, design review, design verification and validation, design transfer, design changes, and design history file your firm was unable to produce any records during the inspection demonstrating that any of your six marketed devices were developed in accordance with Clause 7.3, No design and development files have been maintained for your six products cleared under K223431, K193155, K171750, K012547, K131202, or K901471.
The adequacy of your firm's responses cannot be determined at this time. Your firm initiated CAPA-25026 which commits to performing a DHF Gap Assessment and Reconstruction Program, with a gap assessment completion targeted for (b)(4), and remediation timelines to be established thereafter. However, these items remain in progress as of your July 14, 2026 response. In response to this letter provide documentation for any outstanding corrective actions, including CAPA verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
4. Failure to document procedures to control design and development changes, as required by ISO 13485:2016, Clause 7.3.9.
Specifically, your firm's design control procedure, NRSU-SOP-0018 (Revision 00, Effective Date 05/01/2020), does not include a design change process, and instead references your change control procedure NRSU-SOP-0019. However, NRSU-SOP-0019 does not include a requirement for evaluation of the effect of the changes on constituent parts and product in process or already delivered, including inputs and outputs of risk management and product realization processes, such as design and development verification and validation activities. The following design changes were implemented without the required design and development change records, risk analyses, or documented impact assessments:
• Transition from (b)(4) packaging for the dry product lines in September 2023
• Change in the thickness of the primary packaging film from (b)(4) for both dry product lines in September 2023
• Change of the endotoxin level specification from (b)(4) /mL to (b)(4) /mL for the Liquid Citric Acid Concentrate LC 228 (45x) in March 2025
• Change of the endotoxin level specification from (b)(4) /mL to (b)(4) /mL for (b)(4) Liquid Acidic Concentrate products in March 2025
• Change of the endotoxin level specification from (b)(4) /mL to (b)(4) /mL for the Liquid Acidic Concentrate AC 213 (45x) in April 2025
The adequacy of your firm's responses cannot be determined at this time. Your firm initiated CAPA-25026, which promises retrospective design change impact assessments, initiation of change records, revision to NRSU-SOP-0018 (targeted completion July 31, 2026), and revision to NRSU-SOP-0019 (targeted completion October 31, 2026). In response to this letter provide documentation for any outstanding corrective actions, including CAPA verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
5. Failure to document the requirements for infrastructure needed to achieve conformity to product requirements, prevent product mix-up, and ensure orderly handling of product, as required by ISO 13485:2016, Clause 6.3.
Specifically, your firm has failed to perform maintenance on buildings, workspaces, associated utilities as well as process equipment as required by Clause 6.3(a) and (b). Your procedures NRSU-SOP-0037 'Monitoring Production Systems' (Revision 00, Effective Date 04/13/2020) and NRSU-WI-0040 'Operation of the (b)(4) System' (Revision 00, Effective Date 03/30/2020) govern operation and maintenance of production systems, including the (b)(4) system and bottle filling equipment. During a facility walkthrough on February 13, 2026, our investigator observed multiple pipes leaking in the (b)(4) system. Your firm was unable to provide any documentation related to these leaks, including repairs made, corrective actions taken, or an assessment of the impact of these leaks and repairs on (b)(4) functionality and continued conformity to requirements.
Additionally, the (b)(4) (serial number (b)(4)) is the only instrument used for endotoxin testing including (b)(4) monitoring and final product acceptance. Your firm was unable to provide records demonstrating that the (b)(4) had been changed or that preventive maintenance had ever been performed for this instrument, despite Section 7 of the (b)(4) Operator Manual (Revision A, 2025) requiring (b)(4) replacement and recommended (b)(4) service intervals.
We reviewed your firm's responses and conclude that they are not adequate. Your firm initiated CAPA-25027, which promises implementation of a validated (b)(4) through multiple phases with a target completion date of (b)(4), and an assessment to identify infrastructure with overdue or incomplete maintenance, calibration, or validation activities by (b)(4). However, your firm's responses do not specifically address any corrections or corrective actions planned for the numerous pipe leaks observed, including repairs planned or made or an assessment of the impact on product quality. In addition, provide any documentation for the outstanding corrective actions, including verification of effectiveness. If corrective actions will not be complete, provide a timeframe for completion.
6. Failure to perform calibration or verification of measuring equipment in accordance with documented procedures, as required by ISO 13485:2016, Clause 7.6.
Specifically, your firm's calibration work instruction NRSU-WI-0012 'Calibrations' (Revision 00, Effective Date 04/10/2020) has not been implemented. For example, the (b)(4) (serial number (b)(4)) is the sole instrument used for endotoxin testing for both (b)(4) monitoring and final product release across all product lines. Despite being in use since December 18, 2023, this instrument has never been calibrated, and a calibration schedule has never been established. Furthermore, your firm's work instruction NRSU-WI-0061 (Revision 03, Effective Date 12/03/2025) required (b)(4) calibration reports; however, no reports were provided.
We reviewed your firm's responses and conclude that they are not adequate. While we acknowledge your firm initiated CAPA-25027, which promises implementation of a validated (b)(4) with planned completion by (b)(4), and an infrastructure/maintenance assessment by (b)(4). Your firm has not provided a plan to assess the validity of previous test results obtained using the uncalibrated (b)(4), nor any proposed action for product previously released based on those results. In response to this letter provide any planned evaluation and assessment related to results obtained from this uncalibrated instrument as well as documentation for any other outstanding corrective actions, including verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
7. Failure to document procedures for validation of processes for production and service provisions, as required by ISO 13485:2016, Clause 7.5.6.
Specifically, your firm does not have a documented procedure for validation of processes for production equipment. Your work instruction NRSU-WI-0029 'Validated Systems and Equipment Change' (Revision 00, Effective Date 07/12/2021), was provided as the sole document governing validation. This work instruction does not include all the requirements found within sections a) through g) of Clause 7.5.6.
Additionally, your firm was unable to provide documentation that your (b)(4) (serial number (b)(4)), which is used for endotoxin monitoring of (b)(4) and final product release was validated as the only document provided was a combined Installation, Operational, and Performance Qualification (IOPQ) test results from December 15 and 18, 2023. This document only provides testing results for products with an endotoxin specification of (b)(4) /mL. The following endotoxin testing applications performed routinely by your firm were not included in this IOPQ document and have not been adequately validated:
• Endotoxin testing of all liquid product lines (Liquid Acidic Concentrate, Liquid Citric Acid Concentrate, and Liquid Sodium Bicarbonate Hemodialysis Solution)
• The retesting procedure for endotoxin results found to be out-of-specification.
We reviewed your firm’s responses and conclude that they are not adequate. Your firm initiated CAPA-25002 prior to the current inspection and completed additional performance qualification (PQ) activities for the (b)(4) instrument, which were provided in your July 14, 2026 response. However, this additional PQ (NRSU-VAL-0153 Addendum 1) was limited to your Liquid Acetic Acid products. In response to this letter provide any PQ activities for your other product lines as well as documentation for any other outstanding corrective actions, including verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
8. Failure to document one or more processes for risk management in product realization, as required by ISO 13485:2016, Clause 7.1.
Specifically, your firm has not established a systematic, documented, risk-based approach to identifying, evaluating, and controlling risks associated with product realization processes. For example, your firm does not have policies, procedures, or practices for overall risk management, design risk analysis, or process risk analysis with requirements for continual evaluation and updates. Your firm maintains risk analysis reports for each of its products; however, these reports have not been updated since their initial creation. For example, the risk analysis for MedicaLyte Liquid Bicarbonate (NRSU-RPT-0004, Effective Date 11/13/2020) states that all risks are either negligible or acceptable. Despite an ongoing Class I Recall (Z-1730-2025 / RES 96646) of MedicaLyte Liquid Bicarbonate Concentrate initiated April 11, 2025, due to microbial and fungal contamination, the risk analysis for this product has not been updated since November 13, 2020. Your firm's design control procedure NRSU-SOP-0018 (Revision 00, Effective Date 05/01/2020) requires updates to risk management documentation when known and potential hazards are identified.
The adequacy of your firm's responses cannot be determined at this time. Your firm initiated CAPA-25006 prior to the current inspection with plans to develop a risk management SOP, review and revise existing product and process risk analysis reports once the new SOP is implemented, and train affected staff, with a target completion date of October 31, 2026. In response to this letter provide documentation for any outstanding corrective actions, including verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
9. Failure to monitor supplier performance in meeting requirements for purchased product, as required by ISO 13485:2016, Clause 7.4.1.
Specifically, your firm's supplier management procedure (NRSU-SOP-0022, Revision 04, Effective Date December 2, 2025) documents your process for classification and evaluation of suppliers. However, your firm was unable to provide any records demonstrating that suppliers and services had ever been classified or evaluated in accordance with this procedure.
The adequacy of your firm's responses cannot be determined at this time. Your firm initiated CAPA-25024 prior to the current FDA inspection and has taken immediate corrections by reviewing and updating the approved supplier list (ASL) and increasing supplier oversight until evaluations and/or audits are completed. Corrective actions planned include revision of the supplier management SOP, strengthening supplier audit controls, improvements to the SCAR process, and implementation of quality agreements for all Level I ((b)(4) risk) suppliers, with a target completion date of November 20, 2026. However, several of these items remain in progress as of your July 14, 2026 response. In response to this letter provide documentation for any outstanding corrective actions, including verification of effectiveness. If corrective actions are not complete, provide a timeframe for completion.
Unique Device Identification Violations The inspection revealed that your devices are misbranded within the meaning of section 502(c) of the Act, 21 U.S.C. § 352(c), because a word, statement, or other information required by or under authority of section 519 of the Act, 21 U.S.C. § 360i, to appear on the label or labeling of the devices was not prominently placed thereon with such conspicuousness (as compared with other words, statements, designs, or devices, in the labeling) and in such terms as to render it likely to be read and understood by the ordinary individual under customary conditions of purchase and use. In particular, 21 CFR 801.20(a), which was promulgated under authority of section 519 of the Act, among other provisions, requires, with exceptions not relevant here, that the label of every medical device bears a unique device identifier (UDI) that meets the requirements of 21 CFR Part 801, subpart B, and 21 CFR Part 830. The labels of the Nipro Dry complete dry acid concentrate for hemodialysis devices do not bear such a UDI. For example, the labels for these devices do not include a device identifier within the meaning of 21 CFR 801.3 or 830.3, there is no UDI presented in easily readable plain-text (see 21 CFR 801.40(a)(1)), and there is no UDI presented in machine-readable form that uses automatic identification and data capture (AIDC) technology (see 21 CFR 801.40(a)(2)).
In addition, the Nipro Dry complete dry acid concentrate for hemodialysis is misbranded within the meaning of section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that there was a failure or refusal to furnish any material or information required by or under section 519 of the Act, 21 U.S.C. § 360i, respecting the devices. In particular, 21 CFR 830.300(a) and 830.320(b), both of which were promulgated under section 519 of the Act, among other provisions, require, with exceptions not relevant here, that the labeler of a device submit electronically to FDA’s Global Unique Device Identification Database (GUDID) the information required by 21 CFR Part 830, subpart E, for each version or model required to bear a UDI. Your firm is a “labeler” within the meaning of 21 CFR 830.3 and has not submitted to GUDID any information required by 21 CFR Part 830, subpart E, respecting these devices.
The failure or refusal to furnish any notification or other material or information required by or under section 519 of the Act, 21 U.S.C. § 360i, also constitutes a prohibited act under section 301(q)(1)(B) of the Act, 21 U.S.C. § 331(q)(1)(B). Based on review of the Global Unique Device Identification Database (GUDID) on June 11, 2026, there are no entries for Nipro Dry complete dry acid concentrate for hemodialysis.
A follow up inspection will be required to ensure that corrections and/or corrective actions are adequate.
Your firm should take prompt action to address any violations identified in this letter. Failure to adequately address this matter may result in regulatory action being initiated by the FDA without further notice. These actions include, but are not limited to, seizure, injunction, and civil money penalties.
Other federal agencies may take your compliance with the FD&C Act and its implementing regulations into account when considering the award of federal contracts. Additionally, should FDA determine that you have Quality Management System Regulation violations that are reasonably related to premarket approval applications for Class III devices, such devices will not be approved until the violations have been addressed. Should FDA determine that your devices or facilities do not meet the requirements of the Act, requests for Certificates to Foreign Governments (CFG) may not be granted.
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, as well as an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Your firm’s response should be comprehensive and address any violations included in this Warning Letter. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.
Your firm’s response should be sent via email to Establishment Assessment Team 1 Assistant Director Gina Brackett at CDRHEnforcement@fda.hhs.gov. Please include in the subject line, “CMS Case 732874” when replying. If you have any questions about the contents of this letter, please contact: Sean Moynihan at sean.moynihan@fda.hhs.gov.
Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems. Your firm should investigate and determine the causes of any violations and take prompt actions to address any violations and bring the products into compliance.
Sincerely, /S/
CAPT Cesar A. Perez, PhD, USPHS Acting Deputy Director Office of Regulatory Programs Office of Product Evaluation and Quality Center for Devices and Radiological Health
Cc: Vikram Chandra, Sr. VP QA/RA - Nipro Americas Group Companies, Inc., Vikram.Chandra@nipromed.com