FDA establishment 1000142899, Lewisberry, Pennsylvania, United States. Part of Nipro Renal Solutions USA's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 4
Official action indicated: 1
Warning letters: 1
Recalls since 2016: 3
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
20182018Inspections0
20192019Inspections0vs 2018+0
120202020Inspections1vs 2019+1
20212021Inspections0vs 2020−1
20222022Inspections0vs 2021+0
20232023Inspections0vs 2022+0
20242024Inspections0vs 2023+0
20252025Inspections0vs 2024+0
120262026Inspections1vs 2025+1
20272027 to dateInspections0
Most Cited at This Site
Inspections citing each reference
ISO 13485:2016 Clause 7.5.1: Medical device or batch records not adequate1ISO 13485:2016 Clause 7.5.1: Medical device or batch records not adequateValue1
ISO 13485:2016 Clause 8.3.1: Records of nonconforming product and actions taken not maintained1ISO 13485:2016 Clause 8.3.1: Records of nonconforming product and actions taken not maintainedValue1
ISO 13485:2016 Clause 4.2.3: Medical device file not established and maintained1ISO 13485:2016 Clause 4.2.3: Medical device file not established and maintainedValue1
ISO 13485:2016 Clause 7.5.6: Records of validation of processes and necessary actions not maintained1ISO 13485:2016 Clause 7.5.6: Records of validation of processes and necessary actions not maintainedValue1
ISO 13485:2016 Clause 8.2.4: Internal audits not conducted to ensure QMS meets requirements and is effective1ISO 13485:2016 Clause 8.2.4: Internal audits not conducted to ensure QMS meets requirements and is effectiveValue1
ISO 13485:2016 Clause 4.1.2: Risk-based approach not applied to QMS processes1ISO 13485:2016 Clause 4.1.2: Risk-based approach not applied to QMS processesValue1
ISO 13485:2016 Clause 8.5.2: Corrective action not taken1ISO 13485:2016 Clause 8.5.2: Corrective action not takenValue1
ISO 13485:2016 Clause 8.2.2: Complaint handling records not maintained1ISO 13485:2016 Clause 8.2.2: Complaint handling records not maintainedValue1
Inspections
4 inspections, newest first
Mar 27, 2026
Devices · Compliance: Devices
15 cited references
ISO 13485:2016 Clause 7.5.1. Medical device or batch records not adequate A record for each medical device or batch of medical devices that provides specified traceability and identifies the amount manufactured and amount approved for distribution has not been established and maintained.
ISO 13485:2016 Clause 8.3.1. Records of nonconforming product and actions taken not maintained Records of the nature of nonconforming product and any subsequent action taken, including the evaluation, and any investigation and the rationale for decisions, were not maintained.
ISO 13485:2016 Clause 8.3.1. Nonconforming product not identified and controlled Product which does not conform to product requirements was not identified and controlled to prevent its unintended use or delivery.
ISO 13485:2016 Clause 4.2.3. Medical device file not established and maintained Medical device files have not been adequately established and maintained.
ISO 13485:2016 Clause 7.5.6. Records of validation of processes and necessary actions not maintained Records of the results and conclusion of validation of processes and necessary actions from the validation were not maintained.
ISO 13485:2016 Clause 8.2.4. Internal audits not conducted to ensure QMS meets requirements and is effective Internal audits have not been adequately conducted at planned intervals to determine whether the QMS is effectively implemented and maintained.
ISO 13485:2016 Clause 4.1.2. Risk-based approach not applied to QMS processes A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.
ISO 13485:2016 Clause 8.5.2. Corrective action not taken Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.
ISO 13485:2016 Clause 8.2.2. Complaint handling records not maintained Complaint handling records were not adequately maintained.
ISO 13485:2016 Clause 7.4.1. Supplier performance not monitored Supplier performance in meeting requirements for the purchased product was not monitored.
ISO 13485:2016 Clause 8.2.6. Evidence of conformity to the acceptance criteria not maintained Evidence of conformity to the acceptance criteria for product monitoring and measurement was not adequately maintained.
ISO 13485:2016 Clause 7.3.9. Procedures to control design and development changes not documented Procedures to control design and development changes were not adequately documented.
ISO 13485:2016 Clause 6.4.1. Work environment requirements and procedures not documented Requirements for the work environment, and the procedures to monitor and control the work environment, were not documented where conditions for the work environment can have an adverse effect on product quality.
ISO 13485:2016 Clause 7.3.10. Design and development files not maintained A design and development file was not maintained for each medical device type or medical device family.
ISO 13485:2016 Clause 6.3. Maintenance records not maintained Maintenance records have not been adequately maintained.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been adequately established to control product that does not conform to specified requirements.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been adequately established.
21 CFR 820.100(b). Documentation Corrective and preventive action activities and/or results have not been adequately documented.
21 CFR 820.30(e). Design review - Lack of or inadequate procedures Procedures for design review have not been adequately established.
Voluntary Action Indicated (VAI)
Jan 30, 2013
Devices · Compliance: Devices
None posted
No Action Indicated (NAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports