FDA establishment 3009346095, West Palm Beach, Florida, United States. Part of Technological Medical Advancements's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 3
Official action indicated: 2
Warning letters: 1
Recalls since 2016: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 1040.10(h)(1)(iii). Legible reproductions Legible reproductions of all labels and hazard warnings required by the regulations were not affixed to the laser product, were not provided with the laser product and did not include the information required for positions 1, 2, and 3 of the logotype specified in the CFR.
21 CFR 820.22. Quality Audit/Reaudit - conducted Quality audits have not been performed.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 820.90(b)(1). Procedures for product review,disposition lack of/inadequate Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.198(a). Complaints Complaint files are not adequately maintained.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been adequately established.
21 CFR 1002.13. Failure to submit You did not submit an annual report for products requiring one.
Official Action Indicated (OAI)
Sep 9, 2021
Devices · Compliance: DevicesDevices · Radiation Control and Health Safety Act
14 cited references
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
21 CFR 1040.10(g)(4). Logotype label output information Each laser product does not have affixed a label bearing the DANGER logotype specified in the regulation.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 1040.10(g)(3). Class IV "Danger" label Each Class IV laser product does not have affixed a label bearing the wording "LASER RADIATION--AVOID EYE OR SKIN EXPOSURE TO DIRECT OR SCATTERED RADIATION" in the position specified in the regulation.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 1040.10(h)(2)(i). Reproduction of affixed information The manufacturer did not provide or cause to be provided in all specification sheets and descriptive brochures pertaining to each laser product, a legible reproduction of the class designation, the warning and the information required for positions 1, 2, and 3 of the applicable logotype required b to be affixed to the product.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been established.
21 CFR 820.20(c). Management review - Lack of or inadequate procedures Procedures for management review have not been established.
21 CFR 820.200(a). Servicing - Lack of or inadequate procedures Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
21 CFR 1002.13. Failure to submit You did not submit an annual report for products requiring one.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been adequately established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been implemented.
Voluntary Action Indicated (VAI)
Feb 24, 2020
Devices · Postmarket Assurance: DevicesDevices · Compliance: Devices; Devices · Radiation Control and Health Safety Act
13 cited references
21 CFR 1040.10(h)(2)(i). Reproduction of affixed information The manufacturer did not provide or cause to be provided in all specification sheets and descriptive brochures pertaining to each laser product, a legible reproduction of the class designation, the warning and the information required for positions 1, 2, and 3 of the applicable logotype required b to be affixed to the product.
21 CFR 820.50. Purchasing controls, Lack of or inadequate procedures Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
21 CFR 1040.11(a)(3). Aperture label A label bearing the wording: "Laser aperture". is not affixed to the medical laser product, in close proximity to the aperture through which was emitted accessible laser radiation in excess of the accessible emission limits of Class I.
21 CFR 1010.2(b). Certification other than by tag or label Certification of a product is not in conformance with the alternate means approved by FDA.
21 CFR 820.200(a). Servicing - Lack of or inadequate procedures Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been established.
21 CFR 820.22. Quality audits - Lack of or inadequate procedures Procedures for quality audits have not been established.
21 CFR 820.30(a). Design control - no procedures Procedures for design control have not been established.
21 CFR 1040.10(f)(5)(ii). Emission indicator - Class IIIb & IV An emission indicator which signals sufficiently prior to emission of such radiation, to allow appropriate action to avoid exposure to the laser radiation.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 1002.13. Failure to submit You did not submit an annual report for products requiring one.
21 CFR 820.80(a). Lack of or inadequate procedures - Acceptance activities Procedures for acceptance activities have not been adequately established.
21 CFR 803.17. Lack of Written MDR Procedures Written MDR procedures have not been developed.
21 CFR 820.90(a). Nonconforming product, Lack of or inadequate procedures Procedures have not been established to control product that does not conform to specified requirements.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports