Diowave Laser System
Recall Z-2162-2026, part of FDA recall event 98507. See Technological Medical Advancements's company page.
Technological Medical Advancements LLC started this Class II device recall on Jan 9, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jan 9, 2026
- Status: Ongoing
Software update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthritis, and to temporarily increase local blood circulation, operating at a maximum average power output of 70W, which is beyond the FDA cleared 60W.
- Product
- Diowave Laser System, REF: Diowave 250W
- Quantity
- 16
- Lots and codes
- UDI-DI: G180DIOWAVESTEALTHMAX0. All software versions prior to, but not including V1.81
- Distribution
- US Nationwide distribution in the states of TX, CA, MI, MD, PA, FL, MN, DC, MI.
- Recalling firm
- Technological Medical Advancements LLC, West Palm Beach, FL, United States
- How the firm notified
- Letter
- FDA classified
- May 11, 2026
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports