FDA establishment 3008342021, Dallas, Texas, United States. Part of Ideal Implant's FDA records. Every inspection FDA has posted for the site, with its outcome and cited regulations and its recalls since 2016.
Inspections: 5
Official action indicated: 2
Warning letters: 2
Recalls since 2016: 1
Inspections by fiscal year
FY2018 to FY2027 to date; 1 classified OAI
How to read this
A hatched bar is a year that is not over, so its count covers only part of the year.
21 CFR 820.198(a). Lack of or inadequate complaint procedures Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
21 CFR 820.80(d). Lack of or inadequate final acceptance procedures Procedures for finished device acceptance have not been adequately established.
21 CFR 803.50(a)(1). Report of Death or Serious Injury An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
21 CFR 820.70(i). Software validation for automated processes Software used as part of the quality system has not been validated for its intended use according to an established protocol.
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 820.30(g). Design Validation - Risk analysis not performed/inadequate Risk analysis is inadequate and is incomplete.
21 CFR 820.20(c). Management review - defined interval, sufficient frequency Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency .
21 CFR 820.100(a). Lack of or inadequate procedures Procedures for corrective and preventive action have not been adequately established.
21 CFR 803.50(a)(2). Individual Report of Malfunction An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Voluntary Action Indicated (VAI)
Jul 20, 2012
Devices · Bioresearch Monitoring
3 cited references
21 CFR 812.140(b)(5). Sponsor records of adverse device effects & complaints Records concerning anticipated adverse device effects reports are not all complete.
21 CFR 812.40. Sponsors' general responsibilities For an investigational study, proper monitoring was not ensured.
21 CFR 812.140(b)(2). Sponsor device shipment records inadequate Records of shipment of an investigational device that include the type and quantity of device, date of shipment, and batch number or code mark reports are not all accurate and complete.
Voluntary Action Indicated (VAI)
May 21, 2010
Devices · Bioresearch Monitoring
2 cited references
21 CFR 812.40. Sponsors' general responsibilities For an investigational study, investigators were not provided with the information they need to conduct an investigation properly.
21 CFR 812.35(a)(3)(iv). No 5 day notification of IDE change FDA was not notified within five working days of manufacturing changes that do not constitute a significant change.
Official Action Indicated (OAI)
Warning letters
Letters naming this establishment; each opens the letter
An FEI (FDA Establishment Identifier) names one physical site. Inspections, outcomes and citations come from FDA's Data Dashboard; FDA posts citations only for some inspections. Recalls come from FDA's enforcement reports (iRES), one row per recall even when it covers many lots. A contract site can make products for many companies, so the company shown is the one FDA lists as the site's firm. Form 483 observations are read by Arca from the 483s FDA posts in its FOIA reading room, most of them scans: the first sentence of each observation, with redactions as FDA made them. FDA posts 483s for a minority of inspections.FDA Data Dashboard, warning letters and enforcement reports