IDEAL IMPLANT Structured Breast Implant
Recall Z-2219-2019, part of FDA recall event 83251. See Ideal Implant's company page.
Ideal Implant Incorporated started this Class II device recall on Mar 18, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 3, 2021.
- Classification: Class II
- Started: Mar 18, 2019
- Status: Terminated
The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation.
- Product
- IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
- Quantity
- 1,704 implants
- Lots and codes
- Serial numbers between 0100292 and 0102777.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Country of Canada.
- Recalling firm
- Ideal Implant Incorporated, Dallas, TX, United States
- How the firm notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- FDA classified
- Aug 8, 2019
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports