Arca
Skip to content
Tools
Tools

Cell Marque, SOX-10 (Polyclonal)

Recall Z-3126-2017, part of FDA recall event 78055. See Leica Biosystems Richmond's company page.

Leica Biosystems Richmond Inc. started this Class II device recall on Jul 14, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Aug 2, 2024.

  • Classification: Class II
  • Started: Jul 14, 2017
  • Status: Terminated
  • Products in this event: 7

Why It Was Recalled

Class II

A delay in transit may have contributed to nonconformity as the product was stored outside of defined shipping and handling temperature requirements.

Record

Product
Cell Marque, SOX-10 (Polyclonal), 7 mL, REF PA0813, IVD, This antibody is intended for in vitro diagnostic (IVD) use. SOX-10 Rabbit Polyclonal Primary Antibody is intended for laboratory use in the detection of the SOX-10 protein in formalin-fixed, paraffin-embedded tissue stained in qualitative immunohistochemistry (IHC) testing..
Quantity
37
Lots and codes
Product Code PA0813, Lot No. 47800, Exp. Date 09/24/2019
Distribution
Nationwide Distribution to AR, AZ, CA, CT, FL, GA, IL, IN, KY, LA, MI, MN, NC, NH, NV, NY, OR, PA, SC, TN, TX, VA, and WI.
Recalling firm
Leica Biosystems Richmond Inc., Richmond, IL, United States
How the firm notified
E-Mail
FDA classified
Sep 12, 2017

Other Products in This Recall

Recall event 78055

Leica Biosystems Richmond Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports