Recall Z-2039-2016, part of FDA recall event 74245. See Leica Biosystems Richmond's company page.

Leica Biosystems Richmond Inc. started this Class III device recall on May 2, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 13, 2017.

  • Classification: Class III
  • Started: May 2, 2016
  • Status: Terminated

Why It Was Recalled

Class III

Systems have an improperly activated Windows 7 OS, even though a valid Windows 7 license was associated with each manufactured system. The system appears to be unlicensed. These systems cannot be re-activated using standard Windows licensing procedures and are requiring checks of each potentially affected system to confirm Windows 7 OS is properly activated.

Record

Product
CytoVision Image Analysis and Capture System, an automated cell-locating device. The Cytovision is a rapid metaphase finder, image acquisition and computer aided chromosome analysis system which assists the operator in viewing chromosomes and looking for cellular anomalies. CytoVision enables a qualified Cytogeneticist to rapidly and accurately analyze the chromosome banding pattern.
Quantity
283 systems
Lots and codes
System 23CAP001KXX001 with Serial Number 201021, System 23CAP001KXX001 with Serial Number 201020, System 23CAP001KXX001 with Serial Number 200994, System 23CAP001KFX001 with Serial Number 200752, System 23GSLXXXKFXUPG with Serial Number 201013, System 23GSLXXXKFXUPG with Serial Number 201050, System 23CAP007FXX001 with Serial Number 200930, System 23CAP002KFX001 with Serial Number 201000, System 23CAP001KXX001 with Serial Number 200969, System 23CAP001KFX001 with Serial Number 200942, System 23GSL120KFX001 with Serial Number 200910, System 23CAP001FXX001 with Serial Number 200710, System…
Distribution
Worldwide distribution. US nationwide; Portugal, Spain, Thailand, Italy, China, Romania, Egypt, India, France, Hungary, Mexico, Germany, Norway, Turkey, Pakistan, Malaysia, UK, Luxembourg, Israel, Canada, Switzerland, Taiwan, South Africa, Saudi Arabia, Korea, Singapore and Japan.
Recalling firm
Leica Biosystems Richmond Inc., Richmond, IL, United States
How the firm notified
Letter
FDA classified
Jun 20, 2016

Leica Biosystems Richmond Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.