CIVCO Arm Support, REF 106047
Recall Z-3083-2017, part of FDA recall event 77874. See Med-Tec's company page.
Med Tec Inc started this Class II device recall on Aug 10, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 19, 2024.
- Classification: Class II
- Started: Aug 10, 2017
- Status: Terminated
- Products in this event: 2
Rigid Arm Supports (PN: 106015 and 106047), which were used as part of the Posiboard-2 and Posirest, and were distributed as PN: 106015, 106047, 106521, 106522, 109030, and 109040 between Ocotober 2009 and July 2012 may have a height difference of approximately 17mm.
- Product
- CIVCO Arm Support, REF 106047, distributed as: (a) Rigid Arm Support, Large, Set, Red (PN: 106047) (b) PB2-BDLRL (PN: 109040) (c) Rigid Adj. Arm Support, Large, Set (PN: 106048) (d) PSR2-CF-DLRL-2P Posirest-2 (PN: 106522) The Rigid Arm Supports are utilized on the Posibboard2 and Posirest assemblies, which are intended to position the patients upper extremities during breast and lung diagnostic and radiotherapy treatment.
- Quantity
- 30 units
- Lots and codes
- Codes: M181690, M194500, M203470, M213330, M216050, M142180, M143110, M149600, 0154710, M158550, M165330, M16900, M169330, M142230, M149650, M236810
- Distribution
- Worldwide Distribution - US Distribution to the states of : MI, IL and TX., and to the countries of : Greece, Australia, Czech Republic, Hungary, Norway, Brazil, Italy, Switzerland ,UK, Austria, Slovenia, Croatia, Poland, Germany, Japan, France, Netherland, Sri Lanka, Pakistan, Bangladesh, and Israel
- Recalling firm
- Med Tec Inc, Orange City, IA, United States
- How the firm notified
- FDA classified
- Aug 29, 2017
- CIVCO Arm Support, REF 106015, distributed as: (a) Rigid Arm Support, Standard, Set Red (PN: 106015) (b) Rigid Adj.…Z-3082-2017 · Class II · Aug 10, 2017 · Terminated
Newest first, one product per recall event; search them all
- Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient…Z-2243-2020 · Class II · May 12, 2020 · Terminated
- CIVCO Solstice(TM) SRS Immobilization System, Part Numbers: a) CHS03, b) CHS04 - Product Usage: The Tilting Head…Z-1837-2019 · Class II · Apr 15, 2019 · Terminated
- Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3,…Z-0671-2018 · Class II · Nov 8, 2017 · Terminated
- Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and…Z-1992-2016 · Class II · Mar 31, 2016 · Terminated
- CIVCO Shoulder Retractors, Reusable non-sterile hand grip, REF 302070, packaged 1 set per box, RX. For use in…Z-2763-2015 · Class II · Aug 19, 2015 · Terminated
- Civco, MTSR03 Shoulder Retractor with Wrist Loops Kit with Base, plus two pairs of wrist loops (nylon and sherpa),…Z-0990-2015 · Class II · Dec 11, 2014 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports