Arca
Skip to content
Tools
Tools

Type S Extension For use with Varian

Recall Z-2243-2020, part of FDA recall event 85654. See Med-Tec's company page.

Med Tec Inc started this Class II device recall on May 12, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 25, 2021.

  • Classification: Class II
  • Started: May 12, 2020
  • Status: Terminated

Why It Was Recalled

Class II

Potential for disengagement of the Varian Style Type-S Extension during patient repositioning. The risk associated with the disengagement of the extension could be a patient fall if the patient is not supported by clinical staff.

Record

Product
Type S Extension For use with Varian, CIVCO, Prescription Only - Product Usage: Device is intended for patient positioning.
Quantity
171 units
Lots and codes
Part Number: MTCFHNV UDI: 00841439104156 Lot Numbers: M027990; M505880; M847500; M166160; M505890; M858580; M198170; M505900; M858590; M241650; M505910; M869030; M249180; M511240; M885500; M257630; M530920; M918000; M278760; M541350; M919980; M282330; M553300; M919990; M283380; M553720; M928400; M293670; M553740; M928410; M300750; M596220; M934260; M303010; M596230; M936110; M314410; M616690; M942400; M321880; M616700; M942510; M323560; M649630; M944110; M326580; M649670; M951030; M333420; M649770; M964730; M350700; M715520; M979830; M356130; M715540; M987460; M362290; M743010; M995860;…
Distribution
Worldwide distribution - US Nationwide distribution including in the states of WI, TX, AZ, FL, NE, TN, MA, FL, GA, IN, OR, PA, IL, LA, NH, MS, MI, OH, MN, VA, IA, CA, NC, SC, AK, UT, WA, MO, MT, KS, and WV. The countries of Africa, Colombia, Saudi Arabia, Costa Rica, South Korea, Hungary, Spain, and Italy.
Recalling firm
Med Tec Inc, Orange City, IA, United States
How the firm notified
E-Mail
FDA classified
Jun 2, 2020

Med Tec Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.