Alaris System PC Unit Model 8015

Recall Z-2909-2020, part of FDA recall event 85900. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class I device recall on Aug 4, 2020 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class I
  • Started: Aug 4, 2020
  • Status: Ongoing
  • Products in this event: 2

Why It Was Recalled

Class I

The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.

Record

Product
Alaris System PC Unit Model 8015
Quantity
313740
Lots and codes
All serial numbers manufactured from April 07, 2017 to Present.
Distribution
U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Sep 4, 2020

Other Products in This Recall

Recall event 85900

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.