Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump…
Recall Z-2879-2016, part of FDA recall event 74508. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class II device recall on Aug 8, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 26, 2018.
- Classification: Class II
- Started: Aug 8, 2016
- Status: Terminated
A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.
- Product
- Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.
- Quantity
- 12,000 units
- Lots and codes
- Software version 9.15
- Distribution
- Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Sep 23, 2016
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports