Arca
Skip to content
Tools
Tools

Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump…

Recall Z-2879-2016, part of FDA recall event 74508. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Aug 8, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 26, 2018.

  • Classification: Class II
  • Started: Aug 8, 2016
  • Status: Terminated

Why It Was Recalled

Class II

A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.

Record

Product
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.
Quantity
12,000 units
Lots and codes
Software version 9.15
Distribution
Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Sep 23, 2016

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports