BariAir Therapy System The BariAir Therapy System is a risk management treatment system…
Recall Z-2466-2012, part of FDA recall event 62979. See KCI USA's company page.
KCI USA, Inc. started this Class II device recall on Sep 17, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 29, 2013.
- Classification: Class II
- Started: Sep 17, 2012
- Status: Terminated
KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow
- Product
- BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy.
- Quantity
- 575
- Lots and codes
- Multiple Serial Numbers with Model Numbers: 404000,405500, 405500RF
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
- Recalling firm
- KCI USA, Inc., San Antonio, TX, United States
- How the firm notified
- Letter
- FDA classified
- Sep 27, 2012
- V.A.C. GranuFoam Silver Large Dressing is barrier to bacterial penetration and may help reduce infection. Featuring…Z-0070-2015 · Class II · Sep 18, 2014 · Terminated
- CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50Z-2717-2014 · Class II · May 19, 2014 · Terminated
- RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result…Z-0409-2013 · Class II · Jun 27, 2012 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports