RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary…
Recall Z-0409-2013, part of FDA recall event 62805. See KCI USA's company page.
KCI USA, Inc. started this Class II device recall on Jun 27, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 22, 2013.
- Classification: Class II
- Started: Jun 27, 2012
- Status: Terminated
The firm is initiating a voluntary medical device correction for the RotoRest Delta Therapy System head pack. KCI has received complaints alleging patients experiencing skin breakdown on the head and face due to the positioning of the head pack. In a few instances, the patient received injuries ranging from Stage III to Stage IV pressure ulcers.
- Product
- RotoRest Delta Kinetic Therapy System INDICATIONS: Treatment and prevention of pulmonary complications as a result of immobility, Thoracic or lumbar fracture, Cervical traction, Skeletal traction
- Quantity
- 282 units
- Lots and codes
- Model Number 208030, Serial Numbers: 00157087, 00157091, 00157288, 00157375, 00157405, RRCKOO011, RRHKOO065, RRHKOO074, RRHKOO096, RRHK00109, and RRHKOO117.
- Distribution
- Nationwide Distribution including AL, CA, CO, OK and GA.
- Recalling firm
- KCI USA, Inc., San Antonio, TX, United States
- How the firm notified
- Letter
- FDA classified
- Nov 20, 2012
- V.A.C. GranuFoam Silver Large Dressing is barrier to bacterial penetration and may help reduce infection. Featuring…Z-0070-2015 · Class II · Sep 18, 2014 · Terminated
- CelluTome Epidermal Harvesting System Part number CT-H25 or Part Number CT-H50Z-2717-2014 · Class II · May 19, 2014 · Terminated
- BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or…Z-2466-2012 · Class II · Sep 17, 2012 · Terminated