Arca
Skip to content
Tools
Tools

Alaris EtCO2, Model 8300

Recall Z-2443-2019, part of FDA recall event 83296. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Jul 1, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Sep 9, 2020.

  • Classification: Class II
  • Started: Jul 1, 2019
  • Status: Terminated

Why It Was Recalled

Class II

Calibration; it has been determined that the capnographs may have been calibrated during manufacturing using a lower than specified concentration of CO2

Record

Product
Alaris EtCO2, Model 8300, UDI: 10885403830013, CareFusion for models 11634567 and 8300ADXeN933
Quantity
76 EtCO2 units
Lots and codes
Serial #/Model # 15098573 11634567; 15111502 11634567; 15096360 11634567; 15098699 11634567; 15099681 11634567; 15096411 11634567; 15091045 11634567; 15095827 11634567; 15090452 11634567; 15091016 11634567; 15091061 11634567; 15102989 11634567; 15096388 11634567; 15098536 11634567; 15098577 11634567; 15099809 11634567; 15099880 11634567; 15100031 11634567; 15101644 11634567; 15101802 11634567; 15102453 11634567; 15102532 11634567; 15091069 11634567; 15091118 11634567; 15102694 11634567; 15095732 11634567; 15095813 11634567; 15095837 11634567; 15095838 11634567; 15095745 11634567; 15096145…
Distribution
US: KY, TX CA, IL, MN, NC, IA, SD, KS, DC OUS: Australia
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Aug 30, 2019

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports