Arca
Skip to content
Tools
Tools

ellacor System with Micro-Coring Technology-Indicated for use by medical professionals…

Recall Z-2393-2023, part of FDA recall event 92689. See Cytrellis Biosystems's company page.

Cytrellis Biosystems, Inc. started this Class II device recall on Jun 23, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Jun 23, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring

Record

Product
ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. Model #: Gen 4
Quantity
127 systems
Lots and codes
UDI = B933ELCON010 S/N: SN 0101 0102 0103 0104 0105 0106 0107 0109 0111 0112 0113 0114 0115 0116 0117 0118 0119 0120 0121 0124 0126 0127 0128 0129 0130 0131 0132 0133 0134 0135 0136 0137 0138 0139 0140 0141 0142 0145 0146 0147 0148 0149 0151 0152 0153 0154 0155 0156 0157 0158 0160 0161 0162 0163 0165 0166 0167 0168 0169 0170 0172 0173 0174 0175 0177 0178 0179 0180 0181 0183 0184 0185 0186 0187 0188 0189 0190 0191 0192 0195 0196 0197 0198 0199 0200 0201 0202 0203 0204 0205 0206 0207 0208 0209 0211 0212 0213 0214 0215 0216 0218 0219 0220 0221 0222 0223 0224 0225 0226 0227 0228 0229 0231 0235…
Distribution
Nationwide
Recalling firm
Cytrellis Biosystems, Inc., Woburn, MA, United States
How the firm notified
E-Mail
FDA classified
Aug 10, 2023

Cytrellis Biosystems, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports