Arca
Skip to content
Tools
Tools

Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System…

Recall Z-1648-2023, part of FDA recall event 92187. See Cytrellis Biosystems's company page.

Cytrellis Biosystems, Inc. started this Class II device recall on Apr 7, 2023 (voluntary: firm initiated); FDA lists it as completed.

  • Classification: Class II
  • Started: Apr 7, 2023
  • Status: Completed

Why It Was Recalled

Class II

Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.

Record

Product
Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
Quantity
89 units
Lots and codes
UDI-DI: B933ELCON010 All serial numbers
Distribution
Nationwide
Recalling firm
Cytrellis Biosystems, Inc., Woburn, MA, United States
How the firm notified
E-Mail
FDA classified
May 26, 2023

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports