Over-Head Ceiling Tube Support
Recall Z-2270-2023, part of FDA recall event 92391. See Shimadzu Corporation Medical Systems Division's company page.
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION started this Class II device recall on Apr 10, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Apr 10, 2023
- Status: Ongoing
Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.
- Product
- Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4 units, FluoroSpeed 300 and X1 units
- Quantity
- 1836
- Lots and codes
- All manufactured prior to August 1,2019. REF/UDI-DI/Serial Numbers: CH-200/04540217031436/0162M46304, 0162M48002, 0162M48801, 0162M48902, 0162M48903, 3ZC5C2A21005, 3ZC5C2A4B008, 3ZC5C2A52005, 3ZC5C2A6B011, 3ZC5C2A84014, 3ZC5C2A85003, 3ZC5C2A88019, 3ZC5C2A91005, 3ZC5C2A92004, 0162M46101, 0162M46104, 0162M47505, 0162M47508, 0162M48702, 0162M49006, 0162M49203, 0162M49501, 0162M50007, 0262M46103, 0262M48307, 0262M49709, 0362M46501, 0362M46710, 0362M48206, 0362M49902, 0462M45303, 0462M48001, 0462M48209, 0462M48302, 0462M48902, 3ZC5C2A16003, 3ZC5C2A21006, 3ZC5C2A23017, 3ZC5C2A23019, 3ZC5C2A28011,…
- Distribution
- U.S. Nationwide distribution.
- Recalling firm
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, Kyoto, N/A, Japan
- How the firm notified
- Letter
- FDA classified
- Jul 31, 2023
Newest first, one product per recall event; search them all
- Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems …Z-0499-2026 · Class II · Oct 16, 2025 · Ongoing
- Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System…Z-0347-2027 · Class II · Sep 30, 2025 · Ongoing
- The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic…Z-0317-2025 · Class II · Sep 6, 2024 · Ongoing
- Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the…Z-0139-2025 · Class II · Sep 6, 2024 · Ongoing
- Shimadzu, Trinias, Digital Angiography SystemZ-1292-2024 · Class II · Jan 29, 2024 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports