Trinias unity, Digital Angiography System with Catheterization Table
Recall Z-0139-2025, part of FDA recall event 95406. See Shimadzu Corporation Medical Systems Division's company page.
SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION started this Class II device recall on Sep 6, 2024 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 6, 2024
- Status: Ongoing
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
- Product
- Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.
- Quantity
- 13
- Lots and codes
- UDI-DI: 04540217058327, 04540217049066, 04540217058143 Serial Numbers: 41EFCD7AC001, 41E58C773001, 41EFCD79C001, 41EFCD78B001, 41EFC73A9001, 41EFCD7D4001, 41E58C76B001, 41EFC73D9001, 41EFCD77C001, 41EFC73A5001, 41EFC73C5001, 41EFCD79A001, 41EFCD7A6001.
- Distribution
- US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.
- Recalling firm
- SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION, Kyoto, N/A, Japan
- How the firm notified
- Letter
- FDA classified
- Oct 22, 2024
- Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems …Z-0499-2026 · Class II · Oct 16, 2025 · Ongoing
- Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System…Z-0347-2027 · Class II · Sep 30, 2025 · Ongoing
- The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic…Z-0317-2025 · Class II · Sep 6, 2024 · Ongoing
- Shimadzu, Trinias, Digital Angiography SystemZ-1292-2024 · Class II · Jan 29, 2024 · Ongoing
- Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and…Z-2270-2023 · Class II · Apr 10, 2023 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports