ARTIS is a family of dedicated angiography systems developed for single and biplane…
Recall Z-2241-2024, part of FDA recall event 94892. See Siemens AG/Siemens Healthcare's company page.
Siemens AG/Siemens Healthcare GmbH started this Class II device recall on Jun 26, 2024 (fda mandated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jun 26, 2024
- Status: Ongoing
- Products in this event: 3
A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
- Product
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
- Quantity
- 201 units in total
- Lots and codes
- ARTIS icono biplane 11327600 4056869063317 ARTIS icono ceiling 11328100 4056869295923 ARTIS icono floor 11327700 4056869149325
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
- Recalling firm
- Siemens AG/Siemens Healthcare GmbH, Forchheim, N/A, Germany
- How the firm notified
- Letter
- FDA classified
- Jul 2, 2024
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and…Z-2242-2024 · Class II · Jun 26, 2024 · Ongoing
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and…Z-2240-2024 · Class II · Jun 26, 2024 · Ongoing