Arca
Skip to content
Tools
Tools

ARTIS is a family of dedicated angiography systems developed for single and biplane…

Recall Z-2240-2024, part of FDA recall event 94892. See Siemens AG/Siemens Healthcare's company page.

Siemens AG/Siemens Healthcare GmbH started this Class II device recall on Jun 26, 2024 (fda mandated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Jun 26, 2024
  • Status: Ongoing
  • Products in this event: 3

Why It Was Recalled

Class II

A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

Record

Product
ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Quantity
201 units in total
Lots and codes
ARTIS icono biplane 11327600 4056869063317 ARTIS icono ceiling 11328100 4056869295923 ARTIS icono floor 11327700 4056869149325
Distribution
Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
Recalling firm
Siemens AG/Siemens Healthcare GmbH, Forchheim, N/A, Germany
How the firm notified
Letter
FDA classified
Jul 2, 2024

Other Products in This Recall

Recall event 94892

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc