Tool, Stripper W/BLD
Recall Z-2183-2015, part of FDA recall event 71534. See American Medical Systems Innovation Center - Silicon Valley's company page.
American Medical Systems Innovation Center - Silicon Valley started this Class II device recall on May 28, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 19, 2016.
- Classification: Class II
- Started: May 28, 2015
- Status: Terminated
- Products in this event: 25
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
- Product
- Tool, Stripper W/BLD, 28S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.
- Quantity
- 16,710 total affected d evices
- Lots and codes
- Model 80-10011-002; All product manufactured from April 8, 2012 to April 8, 2015
- Distribution
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
- Recalling firm
- American Medical Systems Innovation Center - Silicon Valley, San Jose, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jul 21, 2015
- Tool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective…Z-2178-2015 · Class II · May 28, 2015 · Terminated
- Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver…Z-2170-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 44S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket…Z-2181-2015 · Class II · May 28, 2015 · Terminated
- Stripper, 0.4mm Disposable EndoStat Fiber, The Fiber Stripper is a tool provided for stripping the protective…Z-2185-2015 · Class II · May 28, 2015 · Terminated
- Fiber Stripper, 150¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off…Z-2164-2015 · Class II · May 28, 2015 · Terminated
- Tool, StripperW/BLD, 10S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket…Z-2173-2015 · Class II · May 28, 2015 · Terminated
- Fiber Stripper, 550¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off…Z-2168-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 12S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket…Z-2175-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 44S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective…Z-2180-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 14S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective…Z-2176-2015 · Class II · May 28, 2015 · Terminated
- 200¿M Lithotripsy Fiber- Reusable; Sure F!ex laser fibers are intended for use in laser-based surgical…Z-2162-2015 · Class II · May 28, 2015 · Terminated
- Sureflex 273 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical…Z-2158-2015 · Class II · May 28, 2015 · Terminated
- 365¿M Lithotripsy Fiber- Reusable; Sure F!ex laser fibers are intended for use in laser-based surgical…Z-2163-2015 · Class II · May 28, 2015 · Terminated
- Fiber Stripper, 365¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off…Z-2167-2015 · Class II · May 28, 2015 · Terminated
- 1X1 Cleaving Tiles; The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on…Z-2171-2015 · Class II · May 28, 2015 · Terminated
- Sureflex 910 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical…Z-2161-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 14S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket…Z-2177-2015 · Class II · May 28, 2015 · Terminated
- Sureflex 550 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical…Z-2160-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 12S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective…Z-2174-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 18S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket…Z-2179-2015 · Class II · May 28, 2015 · Terminated
- Stripper, 0.6mm Disposable EndoStat Fiber, The Fiber Stripper is a tool provided for stripping the protective…Z-2184-2015 · Class II · May 28, 2015 · Terminated
- Stripper, 0.3mm Disposable EndoStat Fiber, The Fiber Stripper is a tool provided for stripping the protective…Z-2186-2015 · Class II · May 28, 2015 · Terminated
- Fiber Stripper, 273¿m, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off…Z-2166-2015 · Class II · May 28, 2015 · Terminated
- Tool, Stripper W/BLD, 28S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective…Z-2182-2015 · Class II · May 28, 2015 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports